Page 30 of 79
Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
ENGLISH
5.1.CHECK THE SELECTED TYPE OF CONE
The icon cone type icon (shape of the beam limiting device) in use should be illuminated
Icon “rectangular” ON:
Indicates that the selected tube head is equipped with rectangular cone
Icon “round” ON:
Indicates that the selected tube head is equipped with round cone
PLEASE NOTE
This selection can be enabled or disabled by means of a specific procedure that has to be performed by an authorized and trained technician
only.
Contact your authorized installer or the local Acteon Customer Service number on page 3 & 4 of this manual for further information.
PLEASE NOTE
After the modification, default exposure values will be automatically changed according to the new settings.
CAUTION
Before making an exposure, be sure that you are using the same cone type and size indicated on the display of the control unit, otherwise all
the pre-programmed exposure settings and relevant dose related information will be completely incorrect!
5.2.CHECK THE SOURCE TO SKIN DISTANCE
Make sure the cone being used matches the type indicated by the cone length icon which should be illuminated on the control panel at this
time (source-skin distance = SSD)
Icon 20 [cm] 8” ON:
Indicates that the selected tube head is equipped with 8” = 20cm (SSD) cone
Icon 30 [cm] 12” ON:
Indicates that the selected tube head is equipped with 12” = 31cm (SSD) cone
PLEASE NOTE
The modification, installation and configuration of a different type of beam limiting device can be enabled or disabled only by authorized
and trained service technician.
Contact your authorized installer or the local Acteon Customer Service number indicated on page 3 & 4 of this manual for further informa-
tion.
PLEASE NOTE
After the modification, default exposure values will be automatically changed.
CAUTION
Before making an exposure, be sure that you areusing the same cone type and size indicated on the display of the control unit otherwise all
the pre-programmed exposure settings and relevant dose related information will be completely incorrect!
It is not contraindicated to use any type of beam limiting other than those provided by de Götzen S.r.l.