6.4 Connecting and disconnecting M+ accessories during use
Do not connect/disconnect the cords when the medical device is switched ON and the footswitch is
pressed.
Do not tighten or loosen the contra-angles when the I-Surge LED micromotor is activated.
Do not tighten or loosen the tips when the handpiece is activated.
6.5 Switching off the medical device
To avoid damaging the I-Surge LED micromotor, it must be rinsed with sterile water less than 30
minutes after use.
1. Remove the bag from the bracket.
2. Remove the irrigation line perforator from the bag.
3. Immerse the irrigation line perforator in a container with distilled water.
4. Press
and run the purge cycle for one minute to rinse the handpiece.
5. Remove the perforator from the container and purge the system until the medical device's irrigation line
is completely empty.
6. Switch off the medical device (O/I switch to O).