6.1.8 Purging the medical device
The medical device must be purged before and after use.
1. Immerse the irrigation line perforator in a container with distilled water.
2. Press
to rinse the irrigation line and the I-Surge LED micromotor or handpiece.
3. Minimum purge time is 1 minute.
6.2 Switching on the medical device
Place the I/O switch to I. The ImplantCenter™ M+ indicates that the device is switched on by an audible signal.
The ImplantCenter™ M+ detects which handpiece cord is connected (Piezotome® or Newtron®) and the
relevant menu will be displayed on the screen.
6.3 Selecting the operating mode
The ImplantCenter™ M+ can be used in any of the following modes:
l
I-Surge: with I-Surge LED micromotor and contra-angles or a straight handpiece
l
Piezotome: with Piezotome® M+ handpiece;
l
Newtron: with Newtron® handpiece.
Switch between modes by pressing the pictograms or using the footswitch as follows.
6.3.1 Common operations using the footswitch
The footswitch is used for common operations.