4 Medical device description
4.1 Control unit
The control unit is compatible with:
1. The I-surge LED rotary technology via an I-surge LED surgical micromotor.
2. Newtron® ultrasonic piezoelectric technology via the surgical Piezotome® M+ handpiece or via the
Newtron® dental handpiece and cord.
The piezo-ultrasonic technology common to both piezoelectric handpieces, called Newtron®, is a patented
system that automatically controls the frequency and power of tips in real time: the Cruise Control® system.
Ultrasonic tips therefore maintain an even efficiency on the work area, regardless of the environment.
Ultrasonic tips are active on hard tissue only, with minimal risks for soft tissues.
The micromotor can be used in combination with either the Piezotome® M+ handpiece, or with the Newtron®
handpiece.
The touch-sensitive interface calls up all programmes stored in the memory and adjusts them before and after
use.
4.2 I-Surge LED micromotor and handpieces
Only SATELEC® handpieces and I-Surge LED micromotor can be connected to the medical device. For further
information, refer to the documentation listed in chapter chapter
.
4.3 Cords
The micromotor cord is only compatible with the I-Surge LED micromotor.
The Newtron® handpiece cord is only compatible with the Newtron® handpiece. It ensures irrigation
circulation and is used to connect the medical device to the Newtron® handpiece.
The Piezotome® M+ cord is only compatible with the Piezotome® M+ handpiece. It ensures circulation and is
used to connect the medical device to the Piezotome® M+ handpiece.
4.4 Inlets
The inlets at the back of the medical device provide suitable ventilation of the control unit and must be kept
clear.
4.5 Switch
The mains switch is used to switch the medical device on (position I) or off (position O).
4.6 Mains connector
The mains connector with its earthing pin is used to connect the device to the electrical network via a
disconnectable mains cord.
4.7 Fuse recess
The mains cord connector holds the recess for two mains fuses designed to protect the medical device in the
event of overvoltage or an internal fault.
4.8 Pump
The medical device is fitted with two peristaltic pumps. A single-use pump is connected to the irrigation line.
User Manual • ImplantCenter™ M+ • J27251 • V1 • (13) • 09/2013 • NO25EN010A - Page
13
/
42
Summary of Contents for ImplantCenter M+
Page 1: ...User Manual ImplantCenter M...
Page 2: ......
Page 5: ...15 Glossary 41...
Page 6: ......
Page 8: ......
Page 26: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 24 42...
Page 28: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 26 42...
Page 40: ...User Manual ImplantCenter M J27251 V1 13 09 2013 NO25EN010A Page 38 42...