Model DT333-0700
4-Wire mV/TC Dual Transmitter w/USB
Acromag, Inc. Tel: 248-295-0880
- 24 -
http://www.acromag.com
- 24 -
https://www.acromag.com
C
alibration…
CALIBRATION
–
FACTORY SETTINGS
You can use the
[Restore Factory Calibration]
button to
restore the transmitter
’s original factory calibration if
you think you made an error during recalibration, have
degraded its performance, or the I/O channel appears
erratic.
You can use the
[Restore to Factory Default]
button to
return the unit to its original factory configuration
settings. This option does not restore calibration, only
configuration. Alternately, this button can be used as a
sanitation tool to restore the unit to its initial
configuration when decommissioning a module.
CALIBRATION
–
COMMUNICATION STATUS
This field displays calibration status messages like “No
Error”, “Transfer Error”, and “Timeout Error” during
calibration. If you encounter a Transfer or Timeout
Error, calibration has not been completed and you may
have to repeat the process.
.
Diagnostics
At this point, after Configuration or Calibration, you
could choose the Diagnostics tab to test module
operation. Clicking
[Start Polling]
of TEST I/O on this
page to trigger the software to periodically read the
input channel(s) and display their value in the fields
below the polling button. Note the simulated lamp next
to the button flashes slowly each time it samples the
inputs. Click
[Stop Polling]
to stop polling the inputs
before moving onto another tab/page.