5.4 Typical wiring method
Description:
(1) Each residual current monitor can be equipped with at most 12 AKH-0.66/L-20 residual current
transformers.
(2) Terminals 1 and 2 of the AIM-R100 residual current monitor and terminals 1 and 2 of the power supply
of ACLP10-24 DC24V power supply shall be connected to AC220V, which can be directly connected to the bus
bar as shown in the figure above, and connected to fuse protection of 6A.
(3) Terminals 13 and 14 of the AIM-R100 residual current monitor are connected to the remote reset button,
which is used for the automatic reset of the instrument to ensure the safety and reliability of the system.
(4) When applied to single-phase branches, AKH 0.66/L-20 needs to pass through two lines of L and N of
each output circuit, and the PE line does not need to be worn in; When applied in three-phase three-wire branch,
the transformer passes through three wires A, B and C. When applied to three-phase four-wire, the transformer
8 9
A C L P 1 0 - 2 4
10 11
13 14
1 2
6 7
1
2
4
5
8
9
11 12
13 14 25 26
20 21 22 23 24
U1 U2
I 2 I 3
I 6
COM2
Z1 Z2
DO1+ DO1-
I 7
I 1
3
6
7
I 4
I 5
10
I 8
COM1
I 12 I 11 I 10 I 9
15 16
A
B
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-R100
Fault
Alarm
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-M100
Acrel
AIM-R100
On
Comm
Medical Residual Current Monitor
Menu/
Test/Esc
L
N
PE