Microplate Washer User Manual
Chapter 11 Maintenance
─
29
─
Chapter 11 Maintenance
Tubing maintenance
The Rinse function can be used to clean out the tubing.
Wash head maintenance and cleaning
Remove the wash head and detach the silicone tubing, remove the sealing
screws at the ends of the wash head, submerge the wash head to into 75%
alcohol to water solution. Use the included cleaning tool to clear out any
debris in the suction needles (long) and dispensing needles (short). Alcohol
can be pushed through the wash head openings using a syringe. Or use a
soft brush to clean the inside of the openings.
Filter cleaning
The circular filter assembly at the bottom of each dip tube in the reagent
bottles can be removed for cleaning. Disconnect the silicone tubing from
the filter assembly, unscrew the top portion of the assembly, and all parts
can be cleaned using alcohol.
Plate Holder
The base of the plate holder has a drain with a liquid collection tube.
Distilled water can be added into the plate holder after use, and the liquid
will be drained out automatically to the waste bottle during a Rinse
procedure. Take care not to allow any solid materials or debris to enter the
drain hole, or the drain could be blocked.
Summary of Contents for MW9600