Microplate Washer User Manual
Chapter 7 Creating a new wash protocol
─
21
─
on the well type selected.
One position: There is one aspiration position in the center of the well.
Two positions: There are two positions, left and right on the well bottom.
Three positions: There are three positions, left middle and right.
Columns
:
Use the number buttons on the keyboard to choose the plate
columns which need to be washed.
When using the 1X8 channel wash head use numbers 1 to 8 to
select the columns for washing.
When using 1X12 channel wash head, use letters A to H to select
the columns for washing.
Pressing the number or letter for a row a second time will cancel
the selection.
Save
:
Press the F3 button to
Save
the new program.
Well type
Aspiration Point Settings
One Position
Two Positions
Three Positions
Flat type
Optional
Default
Optional
U type
Default
Not Optional
Not optional
V type
Default
Not Optional
Optional
C type
Default
Not Optional
Not optional
Summary of Contents for MW9600