Left deviation:
Left deviation means the distance between the starting
position of the dispensing needle to the center of the left row of wells of the
micro plate.
Choose the left deviation function in the calibration menu, then push OK
button. The plate will move so that the wash head needles are aligned
above the left row of wells. Use the UP or DOWN arrow buttons to move
the needles left or right, respectively, to position the needles directly above
the center of the wells. Push OK button to confirm and save the setting.
Right deviation:
Right deviation means the distance from the starting position
of the needles to the center of the right row of wells. Adjust the position using
the up and down arrows so the needles are centered above the right most row
of the microplate.