Microplate Washer User Manual
Chapter 5 Start Up and Software
─
13
─
recommended that wash solution bottle R is filled with distilled water, and this is used for
rinsing after using the microplate washer. It is recommended to use the rinse function
before powering off the instrument if it will not be used for a while.
Select which tubing line is to be rinsed, and rinse time.
Select tubing line from bottle A, B or R
Set the rinse time: 00:01-03:59 (mm:ss), the default time is 2:00.
Rinse
1X8 Calibration
and
1x12 Calibration
:
It’s recommended to calibrate the position of the wash head before use,
although it has been calibrated in the factory with standard micro plates.
Vertical and horizontal positioning can be adjusted in increments of 0.1mm
Choose 1x8 or 1x12 channel wash head, depending on which is used,
press OK to enter into the calibration interface.
Attention! Confirm that the proper wash head is selected
before calibration.
The rinse function can be activated by pressing the “Rinse” button.
Summary of Contents for MW9600
Page 37: ... 32 Note ...