Microplate Washer User Manual
Chapter 5 Start Up and Software
─
12
─
Liquid Sensor:
This allows turning on or off the sensors A, B, and R.
Note: Sensor W for Waste Liquid Level cannot be set to off.
Injection Index
: This is for calibrating the set dispensing volume. The range of
the Injection Index is 0-2.00, and can be adjusted in increments
±
0.01. The
default value is 1.00.
Beeper switch
: The audible beep for indicating an alarm or warning can be
turned on or off.
Purge at Power On:
When the first wash protocol is run
after
the instrument is
powered on, the purge function will run to remove bubbles and prime the
dispensing tubes and wash heads. This can be set for “Single” or “Multiple”.
“Single” setting:
Only the single wash solution that is selected in the protocol
will purge through the tubes and wash head prior to washing.
“Multiple” setting:
All of the active wash solutions (Solutions A, B, R, when
liquid level sensors are enabled) will purge through the dispensing tubing and
wash heads.
3. Maintenance:
This menu includes the Rinse setting, 1X8 channel
calibration and 1X12 channel calibration.
Rinse
: The rinse function can be used for priming the dispensing tubes (removing air
bubbles) and also for rinsing out the dispensing tubes and wash head. It is