58
In response, press the square
MODE
key and read the message
Present the black calibration target to the measuring port. Use the landing pad if possible.
When you have presented the target properly press the square
MODE
key again and the
display will say
Measuring,
followed by three averaging messages. When finished the
display should read
Present the white calibration target to the measuring port. Use the landing pad if possible.
When you have presented the target properly press the square
MODE
key again and the
display will say
Measuring,
then three averaging messages. Then the display should
read
This means that the calibration is complete and the unit has returned to the
Ready-to-
Measure
state. Note that the numbers are an example. No specific calibration process
should return these exact numbers. The two numbers are relative to 10000 and represent
the maximum and minimum change applied. So long as the values are within five percent
of 10000 the process is ok. If the numbers become greater than 10% away from 10000
consult the factory about recalibration services.
>Full Calibrate
About
Measure Black
Ready Hit Mode
Measure White
Ready Hit Mode
###Rdy/Measure
Cal 10050/9960