94MS42 Rev.B1 S/W Ver 2.06
1-5
Cooling Requirements
Minimum pressure and flow rate of the chilled water depends on input
temperature.
!
Cl
If water temperature and water flow do not meet the
following specifications the unit will overheat, one or both
light sources will shut themselves off randomly under heavy
usage and damage to the unit could be the result.
•
AP42CL-8000
C o o ling Requirements
24,000 BTU / hr at id le
65,000 BTU / hr at full
load
Water P ressure
2 6 P S I ( 2 0 G P M )
minimum at 50°F (10°C )
1 0 P S I ( 1 0 G P M )
minimum at 45°F (7°C )