PHSA Laboratories CW Site - Point of Care

Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide

Medical Approval: Dr Benjamin Jung

Medical Approval Date: 27 Dec 2016

Version: 1.2

Implementation Date: 7/2/2019 11:55:59 PM

Folder Name: CW\Point of Care\Whole Blood Glucose

This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS

Page 

8

 of 

12

Allow puncture to Air Dry.  Refer to skin puncture procedures for appropriate 

puncture sites and lancets.

Use second blood drop for testing

Use of microcapillary tube with neonate collections in incubators or difficult 

patients

Performing Patient Test

(Attendees simulate a Patient Test with Levels 1 - 6 Comfort Curve Linearity solutions, 
no self-testing with blood

)

Arterial, line samples, or skin puncture may be used.  Follow reference guides for 
sample collection.  
Neonate or Pediatric Samples may be collected in micro collection pipettes prior to test 
strip application.
It is mandatory to collect Samples from patients in Incubators using the micro collection 
pipettes.
Only the approved micro collection pipette can be used. 

1. Touch ‘Patient Test’

2.

Next Screen displayed will ask for Patient ID.  

NOT CURRENT 

PRACTICE AT PHSA

Pass through this page.

Note:  Use of Patient ID is not activated @ PHSA

3. Confirm selection of Patient Test

Touch  

in lower right corner of screen to bypass this screen.

Confirm the test strip lot as per in the Quality Control Test procedure.

4. Touch the barcode symbol  

 at top right corner of screen to  

enter the lot number of the Test Strip Vial.

Scan the lot number on the test strip vial – hold the meter so
scanner is 4-8 inches in front of vial barcode

5.

Flashing Arrow

” symbol appears.

Insert test strip into test strip port.
Test Strip Port at the top of meter.

Note: Test strip is ALWAYS inserted into the meter BEFORE application of patient 
blood sample!!

6. “Flashing Drop” 

symbol appears.

(up to 5 minutes to apply sample before meter times out – 10 minutes in 
Neonate setting)

Summary of Contents for Inform II

Page 1: ...ing Guide Introduction System Components Charging the AccuChek Inform II on the Base Unit AccuChek Inform II Meter tote Highlights of New AccuChek Inform II Whole Blood Glucose System AccuChek Inform...

Page 2: ...Laser Bar Code Scanner care to be taken not to flash in eyes o Capability for inputting Operator and Patient ID s QC and test strip lot s o No Patient ID s entered at this time Infra red contacts in...

Page 3: ...Blue if used for a charge only base unit Accu Chek Inform II Tote box Each Tote Box is labelled assigned to a Glucose Meter and Base Unit o Storage upright or flat o Contains meter test strip vial QC...

Page 4: ...ow completely An insufficient sample will give an error code Up to 10 minutes to apply sample meter time out No sample re dose Meter must be kept level during sample application especially QC testing...

Page 5: ...oved cleaning wipe DO NOT USE OTHER BRAND eg ACCEL TB wipes are corrosive to the meter touch screen Glucose meters and Base Units are cleaned disinfected After each patient use meter only before repla...

Page 6: ...backspace and delete a character entered incorrectly 2 Screen is set to NUMERIC PHSA 3 Touch symbol to confirm 4 Once entry is complete the main menu appears Entering Operator ID Barcode Scanner 1 To...

Page 7: ...downwards DO NOT allow liquid in test strip port Blue dye in solution can stain clothing Result will appear within 5 seconds 7 Touch RANGE OUT OF RANGE symbol to view the QC range If Result in range t...

Page 8: ...be collected in micro collection pipettes prior to test strip application It is mandatory to collect Samples from patients in Incubators using the micro collection pipettes Only the approved micro co...

Page 9: ...r the capillary puncture Select puncture site Choose appropriate capillary puncture device Puncture site is clean Clean site with soap water if required Disinfect with 2 chlorhexidine in 70 isopropyl...

Page 10: ...is an ERROR CODE Touch Cloud symbol in lower left corner of screen Pre programmed comments appear to select Up to 3 comments can be selected Touch Checkmark in lower right to confirm selection Touch C...

Page 11: ...l assessment of patient important to identify neonatal galactosemia Use a Lab Draw to confirm any unexpected high neonate glucose result Consult with physician During the treatment of parenterally gal...

Page 12: ...ecified Please refer to Lab QMS Page 12 of 12 REVISION APPROVAL LOG Version Revision Type Description of Change Revision Date Technical Approval Medical Approval 1 0 New document 30 Nov 2014 Elvira Ko...

Reviews: