PHSA Laboratories CW Site - Point of Care

Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide

Medical Approval: Dr Benjamin Jung

Medical Approval Date: 27 Dec 2016

Version: 1.2

Implementation Date: 7/2/2019 11:55:59 PM

Folder Name: CW\Point of Care\Whole Blood Glucose

This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS

Page 

4

 of 

12

Test Strip: 

Remove a test strip from the vial –

avoid handling the test strip at FRONT 

EDGE

Replace vial cap immediately after opening.

Print Side up showing arrows in to meter (push strip in to meter till meter 
beeps)

Sample application:

Apply to FRONT EDGE of test strip

Sample requirement:

0.6 uL  –

wipe first drop of blood; use 

second drop of blood

Result Time:

5 seconds

Analytical Measuring Range: 0.6 mmol/L – 33.3 mmol/L

Hematocrit Range:

10 – 65% 

Fill the target window from the FRONT EDGE of the strip (not on top of the 
strip)

Fill the yellow target window completely. An insufficient sample will give an 
error code.

Up to 10 minutes to apply sample – (meter time out)

No sample re-dose

Meter must be kept level during sample application (especially QC testing) –

to avoid fluid entering test strip port

      

Accu-Chek Performa Controls (Quality Control = QC):

Level 1 on vial

(LO - on meter)

Level 2 on vial

(HI - on meter)

Target values are on the Test Strip Vial

Lot # and QC Target Values entered by Point of Care Testing - Lab.

Storage is Room Temperature – away from windows or heaters

Open Vial Expiry is 3 months

– obtained from Point of Care Lab 

as per current practice.

Caution:

QC material has a blue stain that will mark hands and clothing.

Ongoing Operator Competency Minimal Requirements:

Operator ongoing competency – 1 patient test & 2 sets of Quality Control per 
Year

Clinical areas with minimal patient can use the proficiency testing sample as 
the “patient”; or “patient test solution” – both obtained from Point of Care 
Testing – Laboratory.

Operator Lockout -  if minimum requirement is not met annually.

View LearningHub Module at orientation - ACCUCHEK INFORM II Glucose 
Meter Point of Care Testing (Locations:  Children’s Hospital, Women’s 

Summary of Contents for Inform II

Page 1: ...ing Guide Introduction System Components Charging the AccuChek Inform II on the Base Unit AccuChek Inform II Meter tote Highlights of New AccuChek Inform II Whole Blood Glucose System AccuChek Inform...

Page 2: ...Laser Bar Code Scanner care to be taken not to flash in eyes o Capability for inputting Operator and Patient ID s QC and test strip lot s o No Patient ID s entered at this time Infra red contacts in...

Page 3: ...Blue if used for a charge only base unit Accu Chek Inform II Tote box Each Tote Box is labelled assigned to a Glucose Meter and Base Unit o Storage upright or flat o Contains meter test strip vial QC...

Page 4: ...ow completely An insufficient sample will give an error code Up to 10 minutes to apply sample meter time out No sample re dose Meter must be kept level during sample application especially QC testing...

Page 5: ...oved cleaning wipe DO NOT USE OTHER BRAND eg ACCEL TB wipes are corrosive to the meter touch screen Glucose meters and Base Units are cleaned disinfected After each patient use meter only before repla...

Page 6: ...backspace and delete a character entered incorrectly 2 Screen is set to NUMERIC PHSA 3 Touch symbol to confirm 4 Once entry is complete the main menu appears Entering Operator ID Barcode Scanner 1 To...

Page 7: ...downwards DO NOT allow liquid in test strip port Blue dye in solution can stain clothing Result will appear within 5 seconds 7 Touch RANGE OUT OF RANGE symbol to view the QC range If Result in range t...

Page 8: ...be collected in micro collection pipettes prior to test strip application It is mandatory to collect Samples from patients in Incubators using the micro collection pipettes Only the approved micro co...

Page 9: ...r the capillary puncture Select puncture site Choose appropriate capillary puncture device Puncture site is clean Clean site with soap water if required Disinfect with 2 chlorhexidine in 70 isopropyl...

Page 10: ...is an ERROR CODE Touch Cloud symbol in lower left corner of screen Pre programmed comments appear to select Up to 3 comments can be selected Touch Checkmark in lower right to confirm selection Touch C...

Page 11: ...l assessment of patient important to identify neonatal galactosemia Use a Lab Draw to confirm any unexpected high neonate glucose result Consult with physician During the treatment of parenterally gal...

Page 12: ...ecified Please refer to Lab QMS Page 12 of 12 REVISION APPROVAL LOG Version Revision Type Description of Change Revision Date Technical Approval Medical Approval 1 0 New document 30 Nov 2014 Elvira Ko...