PHSA Laboratories CW Site - Point of Care

Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide

Medical Approval: Dr Benjamin Jung

Medical Approval Date: 27 Dec 2016

Version: 1.2

Implementation Date: 7/2/2019 11:55:59 PM

Folder Name: CW\Point of Care\Whole Blood Glucose

This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS

Page 

7

 of 

12

Note:  Lot numbers for QC or Test Strips cannot be manually entered.

5.

“Flashing Arrow” symbol appears.  Insert test strip into test strip port at top 
of meter. 

6.

“Flashing Drop” symbol appears.

Meter is ready for QC sample

Mix the control vials by gently inverting 5 times

Remove cap and wipe vial with tissue

Discard first drop and apply 2nd drop to FRONT EDGE of strip.

Note:  Test strip must ALWAYS be in meter BEFORE application of QC solution 

Position the meter on a flat surface or downwards – DO NOT allow liquid in 
test strip port.
Blue dye in solution can stain clothing
Result will appear within 5 seconds. 

7.

Touch “RANGE”  “OUT OF RANGE” symbol to view the QC range.

If Result in range, touch “Check Mark” to return to Main Menu screen. 

If result is out of reference range for level, ‘Out of Range’ symbol will flash.
 A comment is required when QC is “Out of Range” or if there is an 
“ERROR CODE”.

Touch “Cloud” symbol in lower left corner of screen.

Pre-programmed comments appear to select. 

(Up to 3 comments 

can be selected)

Touch “Checkmark” in lower right to confirm selection.

Touch “Check Mark” again if satisfied with comment to exit from 
screen.

Note: Once exited from QC result/comment screen, new comment cannot be 
added later

Pre-Programmed Comments for Quality Control Testing available to use in Meters

@ PHSA:

QC Procedure Error

Wrong QC Vial

To Repeat QC Test

To Remix QC vial

Repeated QC Test

Open new Strip Vial

Patient Test - Puncture Site  & Sample Collection

Puncture site is clean.  Clean site with soap and water if required.

Disinfectant is 2% Chlorhexidine & 70% isopropyl alcohol

Summary of Contents for Inform II

Page 1: ...ing Guide Introduction System Components Charging the AccuChek Inform II on the Base Unit AccuChek Inform II Meter tote Highlights of New AccuChek Inform II Whole Blood Glucose System AccuChek Inform...

Page 2: ...Laser Bar Code Scanner care to be taken not to flash in eyes o Capability for inputting Operator and Patient ID s QC and test strip lot s o No Patient ID s entered at this time Infra red contacts in...

Page 3: ...Blue if used for a charge only base unit Accu Chek Inform II Tote box Each Tote Box is labelled assigned to a Glucose Meter and Base Unit o Storage upright or flat o Contains meter test strip vial QC...

Page 4: ...ow completely An insufficient sample will give an error code Up to 10 minutes to apply sample meter time out No sample re dose Meter must be kept level during sample application especially QC testing...

Page 5: ...oved cleaning wipe DO NOT USE OTHER BRAND eg ACCEL TB wipes are corrosive to the meter touch screen Glucose meters and Base Units are cleaned disinfected After each patient use meter only before repla...

Page 6: ...backspace and delete a character entered incorrectly 2 Screen is set to NUMERIC PHSA 3 Touch symbol to confirm 4 Once entry is complete the main menu appears Entering Operator ID Barcode Scanner 1 To...

Page 7: ...downwards DO NOT allow liquid in test strip port Blue dye in solution can stain clothing Result will appear within 5 seconds 7 Touch RANGE OUT OF RANGE symbol to view the QC range If Result in range t...

Page 8: ...be collected in micro collection pipettes prior to test strip application It is mandatory to collect Samples from patients in Incubators using the micro collection pipettes Only the approved micro co...

Page 9: ...r the capillary puncture Select puncture site Choose appropriate capillary puncture device Puncture site is clean Clean site with soap water if required Disinfect with 2 chlorhexidine in 70 isopropyl...

Page 10: ...is an ERROR CODE Touch Cloud symbol in lower left corner of screen Pre programmed comments appear to select Up to 3 comments can be selected Touch Checkmark in lower right to confirm selection Touch C...

Page 11: ...l assessment of patient important to identify neonatal galactosemia Use a Lab Draw to confirm any unexpected high neonate glucose result Consult with physician During the treatment of parenterally gal...

Page 12: ...ecified Please refer to Lab QMS Page 12 of 12 REVISION APPROVAL LOG Version Revision Type Description of Change Revision Date Technical Approval Medical Approval 1 0 New document 30 Nov 2014 Elvira Ko...

Reviews: