background image

PHSA Laboratories CW Site - Point of Care

Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide

Medical Approval: Dr Benjamin Jung

Medical Approval Date: 27 Dec 2016

Version: 1.2

Implementation Date: 7/2/2019 11:55:59 PM

Folder Name: CW\Point of Care\Whole Blood Glucose

This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS

Page 

3

 of 

12

Green light indicates connection to a PC or network for data transfer  

Red light indicates trouble or BU is rebooting after a power down.

    Rechargeable Battery Pack:

If fully discharged, battery requires 

6 hours for full charge.  Can be ready 

for use within 20 minutes in an emergency.  Requires Quality Check prior 
to use.  Manufacturer recommendation is 1 hour

Contact Point of Care Technologist is Red light flashing is persistent.

BU light shows: Blue if used for a “charge only” base unit; 

Accu-Chek Inform II Tote box

Each Tote Box is labelled & assigned to a Glucose Meter and 
Base Unit

o

Storage upright or flat

o

Contains meter, test strip vial, QC, capillary tubes, 
Lancets (per PHSA protocol)

o

Disinfectant 2% Chlorhexidine and 70% Isopropyl 
Alcohol swabs

o

Quick Reference guide.

o

Clean Surface of Tote Box regularly

o

After each use when taken in to patient room

Accu-Chek Inform II Test Strips

Test Strip Box and Vial:

Obtain from C&W stores (supply chain top up on the units)

All Lot #’s validated by Point of Care Testing (POCT) in the 
Laboratory

One Lot # per year

50 test strips in a vial

Green box on vials indicates is a “maltose free” (no interference)

Test Strip CODE, Quality Control values and Bar Code is on 
each vial.

Scan Bar Code on the test strip vial with each use

Storage is Room Temperature – away from windows or heaters

Once opened – replace cap immediately

Use all strips to the 

manufacturer’s expiry date

there is no open vial 

expiry date.

Discard plastic Chip Code in box – lab use only.

Summary of Contents for Inform II

Page 1: ...ing Guide Introduction System Components Charging the AccuChek Inform II on the Base Unit AccuChek Inform II Meter tote Highlights of New AccuChek Inform II Whole Blood Glucose System AccuChek Inform...

Page 2: ...Laser Bar Code Scanner care to be taken not to flash in eyes o Capability for inputting Operator and Patient ID s QC and test strip lot s o No Patient ID s entered at this time Infra red contacts in...

Page 3: ...Blue if used for a charge only base unit Accu Chek Inform II Tote box Each Tote Box is labelled assigned to a Glucose Meter and Base Unit o Storage upright or flat o Contains meter test strip vial QC...

Page 4: ...ow completely An insufficient sample will give an error code Up to 10 minutes to apply sample meter time out No sample re dose Meter must be kept level during sample application especially QC testing...

Page 5: ...oved cleaning wipe DO NOT USE OTHER BRAND eg ACCEL TB wipes are corrosive to the meter touch screen Glucose meters and Base Units are cleaned disinfected After each patient use meter only before repla...

Page 6: ...backspace and delete a character entered incorrectly 2 Screen is set to NUMERIC PHSA 3 Touch symbol to confirm 4 Once entry is complete the main menu appears Entering Operator ID Barcode Scanner 1 To...

Page 7: ...downwards DO NOT allow liquid in test strip port Blue dye in solution can stain clothing Result will appear within 5 seconds 7 Touch RANGE OUT OF RANGE symbol to view the QC range If Result in range t...

Page 8: ...be collected in micro collection pipettes prior to test strip application It is mandatory to collect Samples from patients in Incubators using the micro collection pipettes Only the approved micro co...

Page 9: ...r the capillary puncture Select puncture site Choose appropriate capillary puncture device Puncture site is clean Clean site with soap water if required Disinfect with 2 chlorhexidine in 70 isopropyl...

Page 10: ...is an ERROR CODE Touch Cloud symbol in lower left corner of screen Pre programmed comments appear to select Up to 3 comments can be selected Touch Checkmark in lower right to confirm selection Touch C...

Page 11: ...l assessment of patient important to identify neonatal galactosemia Use a Lab Draw to confirm any unexpected high neonate glucose result Consult with physician During the treatment of parenterally gal...

Page 12: ...ecified Please refer to Lab QMS Page 12 of 12 REVISION APPROVAL LOG Version Revision Type Description of Change Revision Date Technical Approval Medical Approval 1 0 New document 30 Nov 2014 Elvira Ko...

Reviews: