background image

PHSA Laboratories CW Site - Point of Care

Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide

Medical Approval: Dr Benjamin Jung

Medical Approval Date: 27 Dec 2016

Version: 1.2

Implementation Date: 7/2/2019 11:55:59 PM

Folder Name: CW\Point of Care\Whole Blood Glucose

This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS

Page 

10

 of 

12

7. Touch  “Range”  Field

  to  view  the  Normal,,  Action,  or  Critical  Results 

Ranges.

Touch “Range , “ Out of Normal Range” , “Out of Critical Range” , “Out of 
Reportable Range” Fields
 to view the PHSA Normal,  Critical Range, Analytical Measuring Ranges for 
Neonates, Pediatrics, or Adults.

A comment is required when Patient Test is “Out of Normal Range” or if there 

is an “ERROR CODE”.

Touch “Cloud” symbol in lower left corner of screen.

Pre-programmed comments appear to select. 

(Up to 3 comments 

can be selected)

Touch “Checkmark” in lower right to confirm selection.

Touch “Check Mark” again if satisfied with comment to exit from 
screen.

Note: Once exited from Patient result/comment screen, a new comment cannot be 
added later

Pre-Programmed Comments when reporting a Patient Test available to use in 

Meters @ PHSA:

Procedure Error

NEO GLU over 5.5 to Chk

To Repeat PT Test

Repeated Test

Repeat with Lab Draw

Doctor Notified

To Use 2

nd

 Drop

Open new Test Strips

8.

Touch “Check Mark” if satisfied with patient result to exit from 

screen.

Review a Patient Test or QC Test Result

9. Touch ‘Review Results – from Main Menu Screen’.

Stored test results are displayed in a sequential list.

Touch    UP  or   DOWN  ARROW    to  scroll up  or down in  the  list.  
Results are grouped by date.

Touch an entry in the list to display the related details.

Summary of Contents for Inform II

Page 1: ...ing Guide Introduction System Components Charging the AccuChek Inform II on the Base Unit AccuChek Inform II Meter tote Highlights of New AccuChek Inform II Whole Blood Glucose System AccuChek Inform...

Page 2: ...Laser Bar Code Scanner care to be taken not to flash in eyes o Capability for inputting Operator and Patient ID s QC and test strip lot s o No Patient ID s entered at this time Infra red contacts in...

Page 3: ...Blue if used for a charge only base unit Accu Chek Inform II Tote box Each Tote Box is labelled assigned to a Glucose Meter and Base Unit o Storage upright or flat o Contains meter test strip vial QC...

Page 4: ...ow completely An insufficient sample will give an error code Up to 10 minutes to apply sample meter time out No sample re dose Meter must be kept level during sample application especially QC testing...

Page 5: ...oved cleaning wipe DO NOT USE OTHER BRAND eg ACCEL TB wipes are corrosive to the meter touch screen Glucose meters and Base Units are cleaned disinfected After each patient use meter only before repla...

Page 6: ...backspace and delete a character entered incorrectly 2 Screen is set to NUMERIC PHSA 3 Touch symbol to confirm 4 Once entry is complete the main menu appears Entering Operator ID Barcode Scanner 1 To...

Page 7: ...downwards DO NOT allow liquid in test strip port Blue dye in solution can stain clothing Result will appear within 5 seconds 7 Touch RANGE OUT OF RANGE symbol to view the QC range If Result in range t...

Page 8: ...be collected in micro collection pipettes prior to test strip application It is mandatory to collect Samples from patients in Incubators using the micro collection pipettes Only the approved micro co...

Page 9: ...r the capillary puncture Select puncture site Choose appropriate capillary puncture device Puncture site is clean Clean site with soap water if required Disinfect with 2 chlorhexidine in 70 isopropyl...

Page 10: ...is an ERROR CODE Touch Cloud symbol in lower left corner of screen Pre programmed comments appear to select Up to 3 comments can be selected Touch Checkmark in lower right to confirm selection Touch C...

Page 11: ...l assessment of patient important to identify neonatal galactosemia Use a Lab Draw to confirm any unexpected high neonate glucose result Consult with physician During the treatment of parenterally gal...

Page 12: ...ecified Please refer to Lab QMS Page 12 of 12 REVISION APPROVAL LOG Version Revision Type Description of Change Revision Date Technical Approval Medical Approval 1 0 New document 30 Nov 2014 Elvira Ko...

Reviews: