4
Important Safety Information
Limitations
• Do not use the meter at high hematocrit levels above 65 % or low hematocrit levels below 10 %.
• Not for use in diagnosis or screening of diabetes mellitus.
• Not for neonatal use.
• Abnormally high concentrations of ascorbic acid (vitamin C) resulting in blood concentrations in excess
of 5 mg/dL may cause inaccurate results. If you are not sure please check with your doctor.
• Do not use the meter system to measure blood glucose in people who are experiencing cardiovascular
collapse (severe shock) or decreased peripheral blood flow.
• Do not use this system during xylose absorption test.
• Not for use on critically ill patients, patients in shock, dehydrated patients, or hyperosmolar patients.
• This system has not been tested at altitudes higher than 10,150 feet.
54988_08100918001_EN.indb 4
9/30/16 5:04 PM
Summary of Contents for Guide
Page 2: ...54988_08100918001_EN indb 2 9 30 16 5 04 PM ...
Page 4: ...54988_08100918001_EN indb 4 9 30 16 5 04 PM ...
Page 20: ...18 Your New System 1 54988_08100918001_EN indb 18 9 30 16 5 04 PM ...
Page 38: ...36 Blood Glucose Tests 3 54988_08100918001_EN indb 36 9 30 16 5 04 PM ...
Page 64: ...62 Meter Settings 5 54988_08100918001_EN indb 62 9 30 16 5 04 PM ...
Page 82: ...80 Review Your Data 6 54988_08100918001_EN indb 80 9 30 16 5 04 PM ...
Page 96: ...94 Wireless Communication and Meter Pairing 7 54988_08100918001_EN indb 94 9 30 16 5 04 PM ...
Page 122: ...120 Technical Information 10 54988_08100918001_EN indb 120 9 30 16 5 04 PM ...
Page 128: ...126 Notes 54988_08100918001_EN indb 126 9 30 16 5 04 PM ...
Page 129: ...127 Notes 54988_08100918001_EN indb 127 9 30 16 5 04 PM ...
Page 130: ...128 Notes 54988_08100918001_EN indb 128 9 30 16 5 04 PM ...
Page 131: ...54988_08100918001_EN indb 129 9 30 16 5 04 PM ...