109
Meter Maintenance and Troubleshooting
9
The time and date setting may
be incorrect.
Make sure the time and date
are correct and adjust, if
necessary.
The test strip may be damaged.
Retest with a new test strip.
Your blood sample may contain
a high level of ascorbate.
Contact your healthcare
professional.
Fluid or foreign material may
be present in the test strip slot.
Remove and reinsert the test
strip or repeat the blood
glucose or control test with a
new test strip. If the problem
persists, contact the Accu‑Chek
Customer Care Service Center
at 1‑800‑858‑8072.
An electronic error has occurred.
Contact the Accu‑Chek Customer Care Service Center at
1‑800‑858‑8072.
54988_08100918001_EN.indb 109
9/30/16 5:04 PM
Summary of Contents for Guide
Page 2: ...54988_08100918001_EN indb 2 9 30 16 5 04 PM ...
Page 4: ...54988_08100918001_EN indb 4 9 30 16 5 04 PM ...
Page 20: ...18 Your New System 1 54988_08100918001_EN indb 18 9 30 16 5 04 PM ...
Page 38: ...36 Blood Glucose Tests 3 54988_08100918001_EN indb 36 9 30 16 5 04 PM ...
Page 64: ...62 Meter Settings 5 54988_08100918001_EN indb 62 9 30 16 5 04 PM ...
Page 82: ...80 Review Your Data 6 54988_08100918001_EN indb 80 9 30 16 5 04 PM ...
Page 96: ...94 Wireless Communication and Meter Pairing 7 54988_08100918001_EN indb 94 9 30 16 5 04 PM ...
Page 122: ...120 Technical Information 10 54988_08100918001_EN indb 120 9 30 16 5 04 PM ...
Page 128: ...126 Notes 54988_08100918001_EN indb 126 9 30 16 5 04 PM ...
Page 129: ...127 Notes 54988_08100918001_EN indb 127 9 30 16 5 04 PM ...
Page 130: ...128 Notes 54988_08100918001_EN indb 128 9 30 16 5 04 PM ...
Page 131: ...54988_08100918001_EN indb 129 9 30 16 5 04 PM ...