Control Tests
2
When to Perform a Control Test
Performing a control test lets you know the meter and test strips are working properly. You should perform a
control test when:
• you open a new test strip box.
• you left the test strip container open.
• you think the test strips are damaged.
• you want to check the meter and test strips.
• the test strips were stored in extreme temperatures, humidity, or both.
• you dropped the meter.
• your test result does not match how you feel.
• you want to check if you are performing the test correctly.
About the Control Solution
• Use only ACCU-CHEK Aviva control solution.
• Close the control solution bottle tightly after use.
• Write the date you open the control solution bottle on the bottle label. The control solution must be
discarded 3 months from the date the control solution bottle was opened (discard date) or on the
Use By
date on the bottle label, whichever comes first.
• Do not use control solution that is past the
Use By
or discard date.
• Refer to the control solution package insert for control solution storage conditions.
• The meter automatically recognizes the difference between the control solution and blood.
• The control results are not displayed in memory.
• The control solution can stain fabric. Remove stains by washing with soap and water.
• Control solution is available for purchase. To order the control solution, talk to your pharmacist or visit
accu-chek.com to order online.
Chapter 2: Control Tests
51801_07021933001.indb 13
5/7/14 8:52 AM
Summary of Contents for Aviva Plus
Page 2: ...51801_07021933001_CVR indb 2 5 7 14 9 12 AM ...
Page 4: ...51801_07021933001 indb 2 5 7 14 8 52 AM ...
Page 8: ...Important Safety Information 6 51801_07021933001 indb 6 5 7 14 8 52 AM ...
Page 14: ...12 Your New System 1 51801_07021933001 indb 12 5 7 14 8 52 AM ...
Page 22: ...20 The ACCU CHEK Softclix Lancing Device 3 51801_07021933001 indb 20 5 7 14 8 52 AM ...
Page 60: ...51801_07021933001_ES indb 2 5 7 14 9 12 AM ...
Page 64: ...Información importante de seguridad 6 51801_07021933001_ES indb 6 5 7 14 9 12 AM ...
Page 70: ...12 Su nuevo sistema 1 51801_07021933001_ES indb 12 5 7 14 9 12 AM ...
Page 78: ...20 El dispositivo de punción ACCU CHEK Softclix 3 51801_07021933001_ES indb 20 5 7 14 9 12 AM ...
Page 115: ...51801_07021933001_CVR indb 3 5 7 14 9 12 AM ...