Meter and Lancing Device Cleaning and Disinfecting
6
Cleaning and Disinfecting the Meter
WARNING
If the meter is being operated by a second person who is providing testing assistance to the user, the
meter and lancing device should be cleaned and disinfected prior to use by the second person.
To clean and disinfect without damaging the meter, follow these procedures carefully.
When to Clean and Disinfect the Meter
• Clean the meter to remove visible dirt or other material prior to disinfecting.
• Clean and disinfect the meter at least once per week and when blood is present on the surface of the
meter.
• Clean and disinfect the meter before allowing anyone else to handle the meter. Do not allow anyone else to
use the meter on themselves for testing purposes.
NOTE
Using cleaning and disinfecting products could result in damage to the meter. If you notice any of the
following signs of deterioration after cleaning and disinfecting your meter, stop using your meter and
contact the ACCU-CHEK Customer Care Service Center at 1-800-858-8072: residue around buttons,
clouding of display, button malfunction, out-of-range control results.
What to Clean and Disinfect
The following parts of the meter should be cleaned and disinfected:
• The area around slots and openings (do not get any moisture in slots or openings)
• The meter display
• The entire meter surface
51801_07021933001.indb 38
5/7/14 8:53 AM
Summary of Contents for Aviva Plus
Page 2: ...51801_07021933001_CVR indb 2 5 7 14 9 12 AM ...
Page 4: ...51801_07021933001 indb 2 5 7 14 8 52 AM ...
Page 8: ...Important Safety Information 6 51801_07021933001 indb 6 5 7 14 8 52 AM ...
Page 14: ...12 Your New System 1 51801_07021933001 indb 12 5 7 14 8 52 AM ...
Page 22: ...20 The ACCU CHEK Softclix Lancing Device 3 51801_07021933001 indb 20 5 7 14 8 52 AM ...
Page 60: ...51801_07021933001_ES indb 2 5 7 14 9 12 AM ...
Page 64: ...Información importante de seguridad 6 51801_07021933001_ES indb 6 5 7 14 9 12 AM ...
Page 70: ...12 Su nuevo sistema 1 51801_07021933001_ES indb 12 5 7 14 9 12 AM ...
Page 78: ...20 El dispositivo de punción ACCU CHEK Softclix 3 51801_07021933001_ES indb 20 5 7 14 9 12 AM ...
Page 115: ...51801_07021933001_CVR indb 3 5 7 14 9 12 AM ...