39
Meter and Lancing Device Cleaning and Disinfecting
6
Wash hands thoroughly
with soap and water.
1
A separate Super
Sani-Cloth should be
used for cleaning and
disinfection. For
disinfecting the meter, get
a new cloth and repeat
step 2, making sure that
the surface stays wet for
2 minutes.
Make sure that no
solution is seen in any
slot or opening.
3
Wash hands thoroughly
with soap and water.
Turn the meter off and
wipe the entire meter
surface with a Super
Sani-Cloth. Carefully wipe
around the test strip slot
and other openings.
Make sure that no liquid
enters any slot or
opening.
2
4
How to Clean and Disinfect the Meter
WARNING
Failure to follow these instructions will damage the meter and stop it from working properly.
• DO NOT clean or disinfect the meter while performing a blood glucose or control test.
• DO NOT get any moisture in slots or openings.
• DO NOT spray anything onto the meter.
• DO NOT immerse the meter in liquid.
• Always use the same product for both cleaning and disinfecting.
51801_07021933001.indb 39
5/7/14 8:53 AM
Summary of Contents for Aviva Plus
Page 2: ...51801_07021933001_CVR indb 2 5 7 14 9 12 AM ...
Page 4: ...51801_07021933001 indb 2 5 7 14 8 52 AM ...
Page 8: ...Important Safety Information 6 51801_07021933001 indb 6 5 7 14 8 52 AM ...
Page 14: ...12 Your New System 1 51801_07021933001 indb 12 5 7 14 8 52 AM ...
Page 22: ...20 The ACCU CHEK Softclix Lancing Device 3 51801_07021933001 indb 20 5 7 14 8 52 AM ...
Page 60: ...51801_07021933001_ES indb 2 5 7 14 9 12 AM ...
Page 64: ...Información importante de seguridad 6 51801_07021933001_ES indb 6 5 7 14 9 12 AM ...
Page 70: ...12 Su nuevo sistema 1 51801_07021933001_ES indb 12 5 7 14 9 12 AM ...
Page 78: ...20 El dispositivo de punción ACCU CHEK Softclix 3 51801_07021933001_ES indb 20 5 7 14 9 12 AM ...
Page 115: ...51801_07021933001_CVR indb 3 5 7 14 9 12 AM ...