Control Tests
2
Understanding Out-of-Range Control Results
WARNING
• The control range applies only to a control result. It only indicates that the test strips and meter are
working properly. Do not use a control result to interpret blood glucose results.
• If the control result is out of range, do not use the meter until you solve the problem. Check this list to
help solve the problem.
Troubleshooting Checks
Action
1. Were the test strips or
control solutions
expired?
Discard the test strips or control solution if either is
past the
Use By
date. If the control solution was
opened more than 3 months ago, discard it. Repeat
the control test with an unexpired test strip and an
unexpired control solution.
2. Did you wipe the tip of the control solution
bottle before use?
Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.
3. Were the caps on the test strip container
and the control solution bottle always
closed tightly?
Replace the test strips or control solution if you think
either was uncapped for some time. Repeat the
control test.
4. Was the test strip used immediately after it
was removed from the test strip container?
Repeat the control test with a new test strip and a
fresh drop of control solution.
5. Were the test strips and control solutions
stored in a cool, dry place?
Repeat the control test with a properly stored test
strip or control solution.
6. Did you follow the directions?
Read Chapter 2, Control Tests, and repeat the control
test.
7. Did you choose the correct control solution
level, either 1 or 2, when you performed
the test?
If you chose the wrong control solution level, you can
still compare the control result to the range printed on
the test strip container.
8. Are you still unsure of the problem?
Contact the ACCU-CHEK Customer Care Service
Center at 1-800-858-8072.
51801_07021933001.indb 16
5/7/14 8:52 AM
Summary of Contents for Aviva Plus
Page 2: ...51801_07021933001_CVR indb 2 5 7 14 9 12 AM ...
Page 4: ...51801_07021933001 indb 2 5 7 14 8 52 AM ...
Page 8: ...Important Safety Information 6 51801_07021933001 indb 6 5 7 14 8 52 AM ...
Page 14: ...12 Your New System 1 51801_07021933001 indb 12 5 7 14 8 52 AM ...
Page 22: ...20 The ACCU CHEK Softclix Lancing Device 3 51801_07021933001 indb 20 5 7 14 8 52 AM ...
Page 60: ...51801_07021933001_ES indb 2 5 7 14 9 12 AM ...
Page 64: ...Información importante de seguridad 6 51801_07021933001_ES indb 6 5 7 14 9 12 AM ...
Page 70: ...12 Su nuevo sistema 1 51801_07021933001_ES indb 12 5 7 14 9 12 AM ...
Page 78: ...20 El dispositivo de punción ACCU CHEK Softclix 3 51801_07021933001_ES indb 20 5 7 14 9 12 AM ...
Page 115: ...51801_07021933001_CVR indb 3 5 7 14 9 12 AM ...