Disinfecting a Device and Accessories for Use by Multiple Clients
Note:
This section applies only when multiple clients will be using the same device. In that
situation, it’s vital to effectively disinfect the device and accessories that have been used by a
client before handling and use by another client.
First Step: Put on protective gloves
Next Step: Clean the device and accessories
1.
Wipe down the device and accessories with disinfectant wipes (Virucidal, Bactericidal,
Pseduomonicidal, Tuberculocidal, Fungicidal)
—
Metrex CaviWipes™. Follow the disinfectant product
manufacturer’s instructions for
cleaning
.
2.
Take a Q-tip with cleaner applied (for example, Windex® or other commercial cleaner) and clean
inside all cracks and crevices. You may want to use a toothbrush or similar brush.
3.
Blow the device off with an air hose, or wipe it dry.
Important! Make sure all soil is removed from the device and accessories. This is vital
before proceeding to disinfecting the equipment.
Next Step: Disinfect the device and accessories
1.
Wipe down the device and accessories with a new disinfectant wipe (Virucidal, Bactericidal,
Pseudomonicidal, Tuberculocidal, Fungicidal)—Metrex CaviWipes. Follow the disinfectant product
manufacturer’s instructions for
disinfecting
.
2.
Make sure to clean inside all cracks and corners, and wipe more than once if necessary to keep the
device wet for a minimum of three (3) minutes.
3.
Allow the device and accessories to dry.
Final Step: Wipe the touch screen
1.
When the device and accessories are dry, wipe down the viewing area (device touch screen,
TouchGuide, Keyguard, etc.) with glass cleaner, so the screen doesn’t discolor.
2.
Allow the equipment to dry.
Accent 800 Hardware Manual