5.37
Impella
®
System for Use During Cardiogenic Shock
Figure 5.42 illustrates the correct configuration of the Impella
®
System components after
transitioning to the standard configuration from the set-up configuration.
Dextrose
Solution
Automated Impella
®
Controller
Purge Cassette
Impella
®
Catheter
Y-Connector
D20
Sodium Chloride (NaCl)
Solution in Pressure Bag
Connector Cable
Figure 5.42 Standard Configuration for Impella
®
System after Transfer from the
Set-up Configuration
Disconnecting the Y
Connector
When you switch to the
standard configuration, you
can simply clamp, disconnect,
and cap the red luer on the Y
connector (as shown in Figure
5.25) or you can disconnect
the Y connector entirely and
connect the yellow luer on
the purge tubing directly to
the yellow check valve on the
Impella
®
Catheter.