7.29
Impella
®
System for Use During Cardiogenic Shock
EMERGENCY SHUTDOWN PROCEDURE
In the unlikely event that the Automated Impella
®
Controller software stops responding, follow
the procedure below to restart the controller without stopping the Impella
®
Catheter.
1.
Press and hold the power switch for 30 seconds.
2.
An “Emergency Shutdown Imminent” alarm will sound at 15 seconds.
3.
The controller will shut down after 30 seconds.
4.
Restart the controller.