5.9
Impella
®
System for Use During Cardiogenic Shock
PRIMING THE PLACEMENT SIGNAL LUMEN (IMPELLA
®
2.5 AND
IMPELLA CP
®
)
Impella Flow
0.0
Purge System
System Power
MUTE
ALARM
EX IT
Impella 2.5 S/N: 123456
30 - 06 - 2011 05:30
100%
L/min
0.0 Max
________________________
0.0 Min
Purge Pressure:
Purge Flow:
Prime Placement Signal Lumen
1. Sq ueeze white f lush v alve until controller b eeps.
Priming Impella Placement Signal Lumen
Figure 5.9 Priming the Placement Signal Lumen
14.
Prime the Impella
®
Catheter placement signal lumen by squeezing the white flush valve
for 10 seconds (see Figure 5.10) until the Automated Impella
®
Controller beeps. The
progress bar shows the progress of the priming.
White Flush
Valve
Figure 5.10 Squeezing the White Flush Valve to Prime the Placement Signal Lumen
15.
When the system detects that the flush solution has reached the target pressure within
the required amount of time, the system will advance to the next screen automatically.
16.
The first step on the next screen prompts you to confirm that fluid is exiting the Impella
®
Catheter.
14.
14.
Prime the Impella
®
Catheter placement signal lumen by squeezing the white flush valve
for 10 seconds (see Figure 5.10) until the Automated Impella
®
Controller beeps. The
progress bar shows the progress of the priming.
5
U
S
IN
G
T
H
E A
U
TO
MA
TED
I
M
P
EL
LA
®
C
O
NT
R
O
LL
E
R
WI
TH
TH
E I
M
P
E
LL
A
®
C
A
TH
ETE
R
Summary of Contents for Impella 2.5
Page 4: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Page 16: ......
Page 22: ......
Page 38: ......
Page 40: ......
Page 108: ......
Page 171: ......
Page 173: ......
Page 181: ......
Page 183: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 210: ...INDEX TBD...