7.12
Instructions for Use & Clinical Reference Manual (US)
IMPELLA
®
2.5 OR IMPELLA CP
®
CATHETER FULLY IN VENTRICLE
If the Impella
®
2.5 or Impella CP
®
Catheter is fully in the ventricle, the following alarm will
appear:
Impella Position In Ventricle
In this situation, the placement screen will appear as shown in Figure 7.7.
Impella Flow
3.5
System Power
MUTE
ALARM
0 . 0
DISPLAY
0 . 0
MENU
PURGE
SYSTEM
Impella CP S/N: 123456
30 - 09 - 2012 05:30
100%
CONTROL
FLOW
A U T O
L/min
4.0 Max
________________________
3.0 Min
Purge System
Purge Pressure: 600 mmHg
Purge Flow: 10.0 ml/hr
Motor
Current
( mA)
10 sec.
Placement
Signal
( mmHg)
90/10
( 37 )
10 sec.
160
0
1000
0
8 08 /7 8 3
( 7 93)
1 .
R ed uce t h e P - lev el t o P - 2 un t i l i m a g i n g i s a v a i la b le.
2 .
R ep osi t i on w i t h i m a g i n g .
3 .
R et ur n t o p r ef er r ed P - lev el a n d con f i r m p osi t i on i n g w i t h i m a g i n g .
Impella Position
in V entricle
Figure 7.7 Impella CP
®
Catheter Fully in Ventricle (similar for Impella
®
2.5)
Actions to take:
1.
Under fluoroscopic guidance, reduce the P-level to P-2 and carefully pull back the
Impella
®
Catheter until the aortic waveform signal is showing.
2.
When you see the aortic waveform signal, pull the catheter back an additional 4 cm.
Yellow Flow Rate
If the Automated Impella
®
Controller detects incorrect or
unknown catheter position,
or if placement monitoring
is suspended, the flow rate
in the flow area appears in
yellow as shown in Figure 7.7.
Summary of Contents for Impella 2.5
Page 4: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Page 16: ......
Page 22: ......
Page 38: ......
Page 40: ......
Page 108: ......
Page 171: ......
Page 173: ......
Page 181: ......
Page 183: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 210: ...INDEX TBD...