8.2
Instructions for Use & Clinical Reference Manual (US)
ALARM DISPLAY
The alarm window is located in the upper left region of the display screen on the front of the
Automated Impella
®
Controller (see Figure 8.1). Alarms are listed in order of priority, with the
highest priority alarm at the top. Up to three alarms may be displayed at one time. The colored
background behind the highest priority alarm will alternate between two shades of that color.
The white panel displayed to the right of the alarm header contains instructions for resolving the
alarm condition. The instructions should be followed in the order given.
Impella Flow
3.5
System Power
0 . 0
DISPLAY
0 . 0
MENU
PURGE
SYSTEM
Impella CP S/N: 123456
30 - 09 - 2012 05:30
100%
CONTROL
FLOW
A U T O
L/min
4.0 Max
________________________
3.0 Min
Purge System
Purge Pressure: 600 mmHg
Purge Flow: 20.0 ml/hr
Impella Position
Wrong
Impella Position Wrong
Suction
1 .
C h eck f i lli n g a n d v olum e st a t us.
2 .
C h eck I m p ella p osi t i on .
3 .
R ed uce P - L ev el.
1 .
R ed uce t h e P - lev el t o P - 2 un t i l i m a g i n g i s a v a i la b le.
2 .
R ep osi t i on w i t h i m a g i n g .
3 .
R et ur n t o p r ef er r ed P - lev el a n d con f i r m p osi t i on i n g w i t h i m a g i n g .
Alarm
Window
Mute Alarm Indicator
Figure 8.1 Alarm Window
MUTE ALARM FUNCTION
Pressing the
MUTE ALARM
button on the upper right of the Automated Impella
®
Controller
display screen will silence the audible alarm indicator for 2 minutes (for red or yellow alarms) or
5 minutes (for white advisory alarms). When an alarm is silenced, the words “MUTE ALARM”
next to the button are replaced by the mute alarm indicator, a crossed-out bell icon (as shown in
Figure 8.1).
ALARM HISTORY SCREEN
The alarm history screen may be accessed through the
MENU
. This screen contains a log of the
alarms that occurred during the case. This log is not maintained when the Automated Impella
®
Controller is powered down or after a power failure. The controller does, however, maintain a
long-term log that is saved after the Automated Impella
®
Controller is powered down or after a
power failure and this information may be downloaded by Abiomed personnel.
Alarms That Resolve On
Their Own
The audible indicator will
shut off if an alarm condition
is resolved before you press
MUTE ALARM
. The visual
message, however, will
continue to be displayed,
with the alarm header on
a gray background, for 20
minutes or until you press
MUTE ALARM
. This allows
you to identify the alarm that
occurred.
Summary of Contents for Impella 2.5
Page 4: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Page 16: ......
Page 22: ......
Page 38: ......
Page 40: ......
Page 108: ......
Page 171: ......
Page 173: ......
Page 181: ......
Page 183: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 210: ...INDEX TBD...