7.15
Impella
®
System for Use During Cardiogenic Shock
IMPELLA
®
5.0 OR LD CATHETER POSITION WRONG
If the Impella
®
5.0 or LD Catheter is fully in the ventricle or fully in the aorta, the following alarm
will appear:
Impella Position Wrong
The Impella
®
5.0 or LD System cannot differentiate between these two conditions. In this
situation, the placement screen will appear as shown in Figure 7.10.
Impella Flow
5.0
System Power
MUTE
ALARM
0 . 0
DISPLAY
0 . 0
MENU
PURGE
SYSTEM
Impella 5.0 S/N: 123456
30 - 06 - 2011 05:30
100%
CONTROL
FLOW
P - 8
L/min
5.2 Max
________________________
4.8 Min
Purge System
Purge Pressure: 600 mmHg
Purge Flow: 20.0 ml/hr
Motor
Current
8 01/8 00
( 8 00)
( mA)
10 sec.
Placement
Signal
( mmHg)
6/2
( 2)
10 sec.
60
-20
1000
0
5.0
Impella Position Wrong
1 .
R ed uce t h e P - lev el t o P - 2 un t i l i m a g i n g i s a v a i la b le.
2 .
R ep osi t i on w i t h i m a g i n g .
3 .
R et ur n t o p r ef er r ed P - lev el a n d con f i r m p osi t i on i n g w i t h i m a g i n g .
Figure 7.10 Impella
®
5.0 Catheter Position Wrong (similar for Impella
®
LD)
Actions to take:
1.
Under fluoroscopic or echocardiographic guidance, determine the catheter position.
2.
Reduce P-level to P-2 and reposition the catheter by either pushing the catheter
forward or pulling it back as needed. Confirm that the placement signal and motor
current are both pulsatile.
3.
If fluoroscopic or echocardiographic imaging is not available, obtain imaging equipment
(fluoroscopy, echocardiography, or chest x-ray) to check the catheter position.
7
PA
TI
EN
T M
A
N
A
G
E
M
EN
T T
O
P
IC
S
Summary of Contents for Impella 2.5
Page 4: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 15: ...2 WARNINGS AND CAUTIONS WARNINGS 2 1 CAUTIONS 2 3...
Page 16: ......
Page 22: ......
Page 38: ......
Page 40: ......
Page 108: ......
Page 171: ......
Page 173: ......
Page 181: ......
Page 183: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 210: ...INDEX TBD...