SOTOX A™ MOBILE ANALYSER | USER GUIDE
4 | EN
1. INTRODUC TION
1.3
SoToxa™ Mobile Analyser specifications
ANALYSER
Dimensions (mm)
L 222 × W 88 × H 61
Weight
Approximately 680g
Detection system
Digital imaging device.
Controls
Cartridge and collector sensors. Cartridge recognition using 2D barcodes for identifying cartridge type, ID, batch and expiry date.
Sensor providing feedback on angle of inclination of analyser. Cartridge heating with temperature control. Internal start-up checks and
process controls during testing. Quality control test cartridges.
Display
LCD colour screen 3.2”
User interface
Five way navigation key, left and right select keys, power button, reset button.
Electrical rating
I5V 3A (via external PSU). The reader is only charged/powered by the supplied PSU and cables.
Languages
English.
Memory
Data storage capacity of over 10,000 tests, the last test result can be accessed on the analyser by the operator.
SD card
SD card connectivity for firmware and cartridge updates and export of diagnostic data.
USB Interface
To communicate with the Software Application Suite if required.
Printer
Connection to Martel MCP7810-323 printer for printing test reports.
OPERATING CONDITIONS
Temperature
5°C to 35°C* This is the operational temperature for running a test.
Note: the rated safe electrical operating temperature of the reader is 0°C to 40°C, but tests must be run between 5°C to 35°C. Test
cartridges must not be at -5°C for more than 10 minutes.
Relative humidity
20 to 80% non-condensing.
Storage temperature
-20°C to 45°C.
Environment
Protect the analyser from direct sunlight, humidity and dust.
POWER SUPPLY
Mains power
Input: 100-240V AC 50-60Hz 1.1A | Output: 15V DC 3A.
Battery power
3.2Ah 11.1V lithium battery
PRINTER
Dimensions (mm)
L 150 × W 80 × H 50.
Weight
363.5g.
Controls
Power button.
* Operation within the temperature range of 5ºC to 35ºC will be tested at 20 to 80% RH, however due to testing limitations it is not possible to accurately control humidity below 10ºC. The
SoToxa Mobile Analyser is operational down to 5˚C at an unquantified relative humidity.
1.4
Transport and unpacking
The SoToxa Mobile Analyser and its accessories are supplied in a convenient carrying case, and all have been fully inspected and tested. Upon receiving the system,
immediately inspect the contents of the case for any transportation damage or missing items (see list of accessories below). If any sign of damage is found, or if any
items are missing, contact Product Support immediately.
The SoToxa Mobile Analyser case is designed to allow the analyser and accessories to be used from within the case, however the analyser can be removed if
required. It is recommended that the analyser is operated and transported in the case, and is secured by the velcro straps to prevent movement when in transit.