What You See on the Display
What You Do
Comments
4-4
Questions?
Refer to Page 10-9 For Technical Support Instructions.
9.
Wait for the monitor to analyze the sample
and display the test result.
The monitor counts down then displays
the test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
10.
Note the test result and whether it
falls within the acceptable range.
If required, scan or manually enter the
comment code, and press
Enter.
The monitor may be enabled to scan or
enter a 1- to 2-digit comment code. If
there is no prompt to enter a comment
code, skip down to step 11.
Control test results can appear as numeric
or PASS/FAIL, depending on how the
test is set up. If numeric, an “X” will appear in
front of a test result that is Out-of-Range.
11.
You may select one of the following:
Press
1
–
Next Level
.
Press
2
–
Repeat Test
.
Press
Menu
to return to the
Menu Mode
menu.
Press
On/Off
to turn off the monitor.
Remove the test strip from the monitor
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
If
1
is selected, the
Next Level
test will
appear. Perform the next level control test by
returning to Step 4.
If
2
is selected, the
Scan or Enter Strip Lot
screen will appear. Return to step 5 to repeat
the test.
Chapter 4 : Control Test
Note:
Following a complete set of Glucose controls, 1-Next Level becomes 1-Exit.
5.5 mmol/L