Monitor Configuration
The Precision Xceed Pro Monitor configuration settings are enabled via the data management system. These settings can
be viewed on the hand-held device, but cannot be changed on it. Settings that are configured through the data
management system are:
Upload interval
Length of time between data uploads.
•
Whether you can continue to test (“allow test” or “warn”) or not (“lockout”).
•
Operator ID
Whether the Operator ID is required, optional, or not used for all tests.
•
Define the Operator ID prompt
•
The number of digits that are acceptable for the Operator ID (can set a minimum and maximum number of digits,
•
between 1 and 30).
With manual entry of the Operator ID, determine whether a check digit is required (modulus 10 or 11).
•
Decide acceptable formats for the Operator ID barcode. The administrator selects one or more of the following:
•
Interleaved 2 (I 2) of 5 (with USS Check Digit, with OPCC Check Digit or without a check digit)
•
Code 128
•
Codabar
•
Code 93
•
Code 39 (with USS Check Digit, with OPCC Check Digit, with an alphanumeric check digit,
•
with full ASCII check digit or without a check digit)
RSS
•
EAN (see Chapter 12 for additional information on barcodes)
•
The monitor can truncate (ignore) the leading digits and/or the last digits of an Operator ID. The monitor can be set to
•
select specific digits of an Operator ID. The number of digits is configured in the data management system.
Patient Tests / Patient ID
Determine whether the Patient ID is required, optional, or not used for patient tests.
•
Define the Patient ID prompt.
•
The number of digits are acceptable for the Patient ID (can set a minimum and maximum number of digits, between
•
1 and 30).
Permission for manual entry of the Patient ID.
•
Determine whether duplicate entry of a manually entered ID is required.
•
With manual entry of the Patient ID, determine whether a check digit is required (modulus 10 or 11).
•
Decide acceptable formats for the Patient ID barcode. The administrator selects one or more of the following:
•
I 2 of 5 (with USS Check Digit, with OPCC Check Digit, or without a check digit)
•
Code 128
•
Codabar
•
Code 93
•
Code 39 (with USS Check Digit, with OPCC Check Digit, with an alphanumeric check digit, with full
•
ASCII check digit or without a check digit)
RSS
•
EAN (see Chapter 12 for additional information on barcodes)
•
The monitor can truncate (ignore) the leading digits and/or the last digits of a Patient ID. The monitor can be set to
•
select specific digits of a Patient ID. The number of digits is configured in the data management system.
Decide whether confirmation of the Patient ID required.
•
Determine how the ID will be confirmed.
•
Decide whether comment code is required, optional, or not used for patient tests that are in the acceptable range for
•
the facility and the test strip.
Decide whether comment code is required, optional, or not used for patient tests that are outside of the acceptable
•
range for the facility and the test strip.
Review Setup
8-1
Chapter 8: Review Setup
8-1