The Precision Xceed Pro Monitor requires little routine maintenance. During testing, the sample remains outside the
monitor, which significantly reduces the possibility of contamination. Follow your facility’s procedures for maintenance
and inspection of the system.
This chapter describes the maintenance tasks for the monitor.
Cleaning the exterior surface of the monitor.
•
Replacing the port protector.
•
Replacing the batteries.
•
Cleaning the Exterior Surface
Cleaning the exterior surface of the Precision Xceed Pro Monitor daily is recommended. Follow your facility’s policies
and procedures for infection control, which may require more frequent cleaning.
The monitor needs to be cleaned only with a damp cloth or a sponge and a mild detergent. It is suggested that the
monitor be turned off while it is being cleaned. Do not immerse or autoclave the monitor or flood it with any liquid.
Acceptable cleaning solutions include alcohol and ammonia based cleaners. Recommended solutions are Sani-Cloth®
HB, Sani-Cloth® Plus, and Super Sani-Cloth®.
Bleach or hydrogen peroxide based cleaners will fade the monitor keypad. Hype-Wipe®, Dispatch® and Virox® 5 wipes
are not recommended as they may fade the monitor keypad. Cleaning solutions not listed have not been tested and may
damage the monitor. Contact Abbott Technical Support or your local Abbott office or distributor for the most up-to-date
information on cleaning solutions.
Maintenance
11-1
Chapter 11: Maintenance of the Precision Xceed Pro Monitor
IMPORTANT
: The Joint Commission recommends cleaning after each patient for infection control.
IMPORTANT
: If a pre-packaged wipe is used for cleaning, excess fluid should be squeezed out to keep
liquid from entering the monitor. Permanent damage could occur.