background image

23

24

Wireless Transmission Range

Your Transmitter and Receiver can maintain a connection at distances 

up to 30 metres (100 feet) in open air without obstructions . Individual 

experiences may vary depending upon the line of sight, walls, trees or 

other obstructions within the area . 
Note: 

Be aware that if you are out of audible range of the Receiver, you will 

not hear any alarms if they occur. The Transmitter has no alarms.

Bathing, Showering and Swimming

The Receiver has exposed ports and should NEVER be submerged in 

liquids, however, your Sensor and Transmitter are water resistant .
You can wear the Transmitter/Sensor Unit while bathing or showering . 

You may also swim while wearing the Transmitter/Sensor Unit . Do NOT 

go deeper than 1 metre (3 feet) .

Sleeping

The Transmitter/Sensor Unit should not interfere with your normal 

sleeping patterns . As you get ready to go to sleep, place the Receiver 

nearby to hear any alarms that may occur . It is recommended to charge 

your Receiver overnight to avoid data interruption .

Visibility

The Receiver Display may be less visible in direct sunlight .

Troubleshooting

For troubleshooting information, see Section 11 in the Expert Guide . 

Call your healthcare professional or Customer Care for additional 

information . 

Living with Your FreeStyle Navigator II System

Reconnecting to a Sensor if Signal is Interrupted

If your Receiver and Transmitter become disconnected for any reason 

(as indicated by the 

  icon), you can reconnect them by following the 

steps below .
1.  Make sure that the Transmitter is firmly attached to the Sensor .
2.  The Receiver will automatically try to reconnect with the Transmitter 

within the first 30 minutes of being disconnected . If it has been 

more than 30 minutes, or if you want to re-establish connection 

immediately, select Menu 

 Connect to Sensor .

3.  Hold the Receiver directly on top of the Transmitter . When 

reconnection is successful, you will see the 

  icon on the Home 

screen .

Your FreeStyle Navigator II System can be used during a wide variety of 

indoor and outdoor activities . Depending on the activity, there are some 

considerations to be aware of:

Alarm Snooze Setup

The FreeStyle Navigator II System includes a Low Glucose Snooze and High Glucose Snooze feature that allows you to set how frequently a Low 

Glucose or High Glucose Alarm repeats after you have already cleared the alarm .

Accept

SNOOZE SETUP

Low Glucose =  30  min

High Glucose =  60  min

Menu 

 Alarms 

 Snooze Setup

Low Glucose

15 - 60 minutes

How soon you want to temporarily disable Low Glucose Alarms after you have cleared one

High Glucose

15 - 240 minutes

How soon you want to temporarily disable High Glucose Alarms after you have cleared one

Note: 

If it typically takes one to two hours for your glucose levels to fall after a correction bolus or injection, you may want to set the 

High Glucose snooze time to 60 min or 120 min to avoid unnecessary alarms.

Responding to Alarms

The title of alarm messages describes the cause of the alarm . Details on what the alarm means and what you can do are provided in the Receiver 

message . See Section 8 in the Expert Guide for a table listing other alarm messages that you may see on your Receiver . The possible cause(s) and 

recommended action for each alarm is provided .
Note: 

If you have tried the recommended actions and need additional help or have questions, call your healthcare team or Customer Care.

Softkey

Buttons

OK

CAL ERROR

Check your BG in 15
minutes to calibrate.

If You Wish To . . . 

Action

 .  .  . temporarily mute the alarm

Press the Right Softkey button to temporarily mute the alarm . The alarm will repeat every 
5 minutes until it is cleared .

 .  .  . clear the alarm

Press the Right Softkey button to turn on the display, then press the Left Softkey button 
to clear the alarm . Note that the receiver alarm may sound again after 15 minutes if your 
glucose remains above or below your High or Low Glucose Thresholds .

If an alarm occurs while the Receiver is displaying a screen besides the Home screen, you receive a flashing Pending Alarm symbol 

 in the upper 

right corner of your Receiver screen . You also receive a Pending Alarm symbol if multiple alarms occur at the same time .

ART22683-101_rev-B.indd   23-24

1/7/15   11:43 AM

Summary of Contents for FreeStyle Navigator II

Page 1: ...Continuous Glucose Monitoring System User s Manual ART22683 101_rev B indd 1 1 7 15 11 43 AM ...

Page 2: ...ng Glucose Targets and Alarms 18 Glucose Targets 18 Glucose Alarms Settings 19 Glucose Alarms 20 Alarm Tones 20 Day Night Glucose Alarms 21 Audio Vibrate Settings 22 Mute Un Mute Alarms 22 Alarm Snooze Setup 23 Responding to Alarms 23 Troubleshooting 24 Living with Your FreeStyle Navigator II System 24 Reconnecting to a Sensor if Signal is Interrupted 24 Wireless Transmission Range 24 Bathing Show...

Page 3: ...inually measuring interstitial fluid glucose levels in people age 6 and older with diabetes mellitus The indication for children age 6 17 is limited to those who are supervised by a caregiver who is at least 18 years of age The caregiver is responsible for managing or assisting the child to manage the FreeStyle Navigator II System and also for interpreting or assisting the child to interpret FreeS...

Page 4: ...le use component that attaches to your skin with an adhesive pad Designed to hold the Transmitter and Sensor on your body for up to 5 days The combination of the Sensor Support Mount with Sensor and the Transmitter that is worn on your body is known as the Transmitter Sensor Unit System Related Information The FreeStyle Navigator II System is designed as a complete system Use only the FreeStyle Na...

Page 5: ... on display Temporarily mute an alarm Blood Glucose Meter Test Strip Port and Light USB Port Rotate wheel to scroll among menu options or data Press in to select menu options and to confirm edits ScrollWheel Reset Button Back Button Press and hold to return to Home screen Cancel selection Move to a previous screen without saving changes Note Pressure or force applied to the screen of the Receiver ...

Page 6: ...e may be shortened if you activate the display frequently The number of alarms you receive may also shorten battery life Upon disconnecting from the Charging Cable the Receiver completes a sequence of self checks If the display is turned on the Receiver emits tones vibrates and flashes screen checks during this time If the Receiver Charging Cable Receiver and A C Wall Charger can detect that any p...

Page 7: ...ld occur or the Sensor Support Mount may not stick to the site 2 Remove the Sensor Inserter and Sensor Support Mount from their sterile packages Save the package of the Sensor Inserter it contains the Sensor Code number You need to enter this code number into the Receiver at a later step Note Make note of the Sensor Code found on the Sensor Inserter packaging You will need to enter this code after...

Page 8: ...has been removed and the Insertion Buttons are pressed a needle quickly goes just under your skin to place the Sensor Do NOT press the buttons until you are ready to insert the Sensor 7 Hold the black part of the Sensor Inserter avoiding the blue release tabs Press down firmly on both grey Insertion Buttons on the top of the Sensor Inserter Try not to press the Sensor Inserter assembly down into t...

Page 9: ...r serial number the Sensor Code screen is displayed 4 Using the Scroll Wheel enter the 3 digit Sensor Code from the Sensor Inserter packaging CAUTION The code numbers MUST match to ensure accurate glucose results Note Once the Sensor Code has been entered and you have pressed Accept you cannot change the Sensor Code number Be careful If you enter the code incorrectly you have to replace the Sensor...

Page 10: ...rglycaemia GLUCOSE READING Control Solution No Home Low GLUCOSE READING Control Solution No Home High Notes Use a new sterile lancet each time NEVER use a lancet or lancing device for more than one person Follow instructions for your lancing device Apply blood sample and wait for results 1 Gently apply blood to ONLY ONE sample area of the test strip Only a very tiny drop of blood is needed Note Do...

Page 11: ...ting the Transmitter within approximately 1 minute you may get a screen which reads TX DETACHED Transmitter has detached from Sensor CGM is not available Replace the Sensor to continue CGM If this screen is displayed press OK If you don t separate the Transmitter from the Sensor Support Mount the Receiver displays a screen which reads Did you remove the Sensor Select Yes to end CGM Press Yes 3 Dis...

Page 12: ...Data Loss System Beep Pulse Harp Zing Delight Rhythm Vibe vibrate Off Note Exercise caution when turning off alarms For example if you turn off the Low Glucose Alarms you do NOT receive text audible nor vibratory notification for low glucose events Glucose Alarms Settings Type What It Is Settings Available OK LOW GLUCOSE Low glucose below 75 mg dL 74mg dL The low glucose alarm notifies you when yo...

Page 13: ...s for more than 12 hours you must select Menu Alarms Tones and turn off alarms one at a time Day Night Glucose Alarms GLUCOSE ALARMS Night Start 21 00 Day Start 07 00 Next Menu Alarms Glucose Alarms Glucose Alarms Options Comments Day Start 00 00 24 00 Day and Night Alarm settings allow you to define different glucose alarm thresholds for different times of the day Night Start 00 00 24 00 Day and ...

Page 14: ... wide variety of indoor and outdoor activities Depending on the activity there are some considerations to be aware of Alarm Snooze Setup The FreeStyle Navigator II System includes a Low Glucose Snooze and High Glucose Snooze feature that allows you to set how frequently a Low Glucose or High Glucose Alarm repeats after you have already cleared the alarm Accept SNOOZE SETUP Low Glucose 30 min High ...

Page 15: ... 2 Wash the skin in mild soap and water and dry thoroughly 3 Check to make sure the inside of the skin is completely dry before putting it back on the Receiver Travelling Check with local authorities prior to departure as rules and regulations may change without notice Follow these guidelines when travelling Notify security personnel of the presence of the device when going through security system...

Page 16: ...000 feet Operating Pressure 1 0 bar 14 7 psia sea level to 0 7 bar 10 1 psia 3 048 metres 10 000 feet Sensor Life Up to 5 days Sensor Operating Skin Surface Temperature 25 ºC to 40 ºC 77 ºF to 104 ºF Glucose Result Range 20 to 500 mg dL 1 1 to 27 8 mmol L Receiver Memory 60 days of normal use including continuous glucose readings stored every 10 minutes and daily blood glucose readings Transmitter...

Page 17: ...y charge An icon without a plug indicates that the battery is not currently being charged Charging An icon with a plug indicates the battery is charging Charged If the entire icon is green this indicates the Receiver is completely charged Audio Vibrate Audio and Vibrate The audio and vibratory settings for your Receiver The settings apply to glucose data loss and system alarms but not to your Prog...

Page 18: ...facturer s declaration electromagnetic immunity The System is intended for use in the electromagnetic environment specified below The customer or the user of the System should assure that it is used in such an environment IMMUNITY test IEC 60601 test level Compliance Level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Flo...

Page 19: ... by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures object...

Page 20: ...o frequency of transmitter m 150 kHz to 80 MHz 1 2 d 80 MHz to 800 MHz 1 2 d 800 MHz to 2 5 GHz 2 3 d 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the max...

Page 21: ...Abbott Diabetes Care Inc 1360 South Loop Road Alameda CA 94502 USA PRT22683 101 in various jurisdictions 2011 2013 ART22683 101 Rev B 01 15 ART22683 101_rev B indd 37 1 7 15 11 43 AM ...

Reviews: