Electromagnetic Compatibility
• The System needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in
this manual .
• Portable and mobile RF communications equipment can affect the System .
• The use of accessories, transducers and cables other than those specified by Abbott Diabetes Care may result in increased EMISSIONS or decreased
IMMUNITY of the System .
• The System should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the System should be
observed to verify normal operation in the configuration in which it will be used .
Guidance and manufacturer’s declaration – electromagnetic emissions
The System is intended for use in the electromagnetic environment specified below . The customer or the user of the System should assure that it is used
in such an environment .
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The System uses RF energy only for its internal function . Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment .
RF emissions
CISPR 11
Class B
The System is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low voltage power supply network that supplies buildings used for domestic
purposes .
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The System is intended for use in the electromagnetic environment specified below . The customer or the user of the System should assure that it is
used in such an environment .
IMMUNITY
test
IEC 60601
test level
Compliance
Level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile . If floors are covered
with synthetic material, the relative humidity should be at least 30% .
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that of a typical domestic, commercial or
hospital environment .
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical domestic, commercial or
hospital environment .
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
) for
0 .5 cycle
40%
U
T
(60% dip in
U
T
) for
5 cycles
70%
U
T
(30% dip in
U
T
) for
25 cycles
<5%
U
T
(>95% dip in
U
T
) for
5 seconds
<5%
U
T
(>95% dip in
U
T
) for
0 .5 cycle
40%
U
T
(60% dip in
U
T
) for
5 cycles
70%
U
T
(30% dip in
U
T
) for
25 cycles
<5%
U
T
(>95% dip in
U
T
) for
5 seconds
Mains power quality should be that of a typical domestic, commercial
or hospital environment . If the user of the System requires continued
operation during power mains interruptions, it is recommended that the
System be powered from an uninterruptible power supply or a battery .
Note:
U
T
is the a.c. mains voltage prior to application of the test level.