The Receiver displays your blood glucose results when the check is 

complete . If Progress Tones are on, you also hear a chime-like Progress 

Tone when results appear .
This screen displays your blood glucose result . Press Home to return to 

the Home screen .
Note: 

Measurements are accurate between 5 ° to 40 °C (40 ° to 104 °F). If the 

temperature is out of range, the Receiver displays the

   

symbol with your 

glucose reading.

Discard the test strip and lancet

•  Test strips may be used only once . Discard used test strips in a 

sealable container such as a sharps container to avoid contact with 

biohazards .

Your blood glucose results
Work with your healthcare team to determine the low and high values 

for your blood glucose target range . If you get results below or above 

your target range and do NOT have symptoms of hypoglycaemia or 

hyperglycaemia, repeat the test . If you have symptoms or continue to 

get results that are low or high, follow the treatment recommended by 

your healthcare team .

CAUTION: Low or high glucose readings can indicate a potentially 

serious medical condition .

If Low appears on your Receiver, your result is lower than 20 mg/dL 

(1 .1 mmol/L) . If you have symptoms of low blood glucose (for 

example: weakness, sweating, nervousness, headache or confusion), 

follow your healthcare team’s recommendations for 

treating severe hypoglycaemia . If you do not have 

any symptoms, wash and dry your hands and repeat 

the blood glucose check on your finger with a new 

test strip . If you receive a second Low result, follow 

your healthcare team’s recommendations for 

treating severe hypoglycaemia .

If High appears on your Receiver, your result is higher than 500 mg/dL 

(27 .8 mmol/L) . If you have symptoms of high blood glucose (for 

example: fatigue, thirst, excess urination or blurry 

vision), follow your healthcare team’s 

recommendations for treating hyperglycaemia . If 

you do not have any symptoms, wash and dry your 

hands and repeat the blood glucose check on your 

finger with a new test strip . If you receive a second 

High result, follow your healthcare team’s 

recommendations for treating severe 

hyperglycaemia .

GLUCOSE READING

Control Solution =  No

Home

Low

GLUCOSE READING

Control Solution =  No

Home

High

Notes: 

•  Use a new sterile lancet each time. NEVER use a lancet or lancing 

device for more than one person.

•  Follow instructions for your lancing device.

Apply blood sample and wait for results
1.  
Gently apply blood to ONLY ONE 

sample area of the test strip . Only a 

very tiny drop of blood is needed .

Note: 

Do NOT apply blood to both 

sample areas of the test strip. If this 
happens, discard the test strip.

2.  Watch the Receiver Display . When enough blood is applied, the 

Receiver shows a circle of 4 arrows 

  while it measures the blood 

glucose . If Progress Tones are on, you also 

hear a two-note Progress Tone .

Note: 

If needed, you have up to 60 seconds 

to add blood to the same sample area of the 
test strip.

Test Strip 

Sample Area

GLUCOSE READING

Control Solution =  No

Home

101

mg/dL

101 mg/dL = 5.6 mmol/L

Calibrating

Calibrating is the process the FreeStyle Navigator II System uses to 

match your interstitial fluid glucose readings with your blood glucose 

readings . You need to calibrate the system by checking your blood 

glucose at approximately 1, 2, 10, 24 and 72 hours after Sensor 

insertion .

•  When the system is ready to be calibrated, it 

prompts you to “Check your BG to calibrate” 

and emits an audible alarm based on your 

settings . You will also see a blood drop 

  icon  on the Home screen . To calibrate the 

system, perform a blood glucose check .

Note: 

If your system alarms are turned off, you 

will not see the “Check your BG to calibrate” 
screen or hear audible alarms.

•  See the Expert Guide for more information 

on performing and troubleshooting 

calibrations .

•  There is a time window in which each 

calibration needs to be completed called a 

grace period . If you do not calibrate before 

the grace period ends, the continuous 

glucose results are no longer displayed and the glucose alarms 

become inactive . Your Sensor remains active . If you missed the grace 

period, simply perform a blood glucose check at any time you see 

the blood drop    icon to continue the calibration process; this 

resumes continuous glucose monitoring .

OK

TIME TO CAL

Check your BG to
calibrate.

117

mg/dL

320

180

40

12 May 2012

Graph

14:30

Menu

117 mg/dL = 6.5 mmol/L

mmol/L

18

10

2

Blood Drop Icon

ART22683-101_rev-B.indd   15-16

1/7/15   11:43 AM

Summary of Contents for FreeStyle Navigator II

Page 1: ...Continuous Glucose Monitoring System User s Manual ART22683 101_rev B indd 1 1 7 15 11 43 AM ...

Page 2: ...ng Glucose Targets and Alarms 18 Glucose Targets 18 Glucose Alarms Settings 19 Glucose Alarms 20 Alarm Tones 20 Day Night Glucose Alarms 21 Audio Vibrate Settings 22 Mute Un Mute Alarms 22 Alarm Snooze Setup 23 Responding to Alarms 23 Troubleshooting 24 Living with Your FreeStyle Navigator II System 24 Reconnecting to a Sensor if Signal is Interrupted 24 Wireless Transmission Range 24 Bathing Show...

Page 3: ...inually measuring interstitial fluid glucose levels in people age 6 and older with diabetes mellitus The indication for children age 6 17 is limited to those who are supervised by a caregiver who is at least 18 years of age The caregiver is responsible for managing or assisting the child to manage the FreeStyle Navigator II System and also for interpreting or assisting the child to interpret FreeS...

Page 4: ...le use component that attaches to your skin with an adhesive pad Designed to hold the Transmitter and Sensor on your body for up to 5 days The combination of the Sensor Support Mount with Sensor and the Transmitter that is worn on your body is known as the Transmitter Sensor Unit System Related Information The FreeStyle Navigator II System is designed as a complete system Use only the FreeStyle Na...

Page 5: ... on display Temporarily mute an alarm Blood Glucose Meter Test Strip Port and Light USB Port Rotate wheel to scroll among menu options or data Press in to select menu options and to confirm edits ScrollWheel Reset Button Back Button Press and hold to return to Home screen Cancel selection Move to a previous screen without saving changes Note Pressure or force applied to the screen of the Receiver ...

Page 6: ...e may be shortened if you activate the display frequently The number of alarms you receive may also shorten battery life Upon disconnecting from the Charging Cable the Receiver completes a sequence of self checks If the display is turned on the Receiver emits tones vibrates and flashes screen checks during this time If the Receiver Charging Cable Receiver and A C Wall Charger can detect that any p...

Page 7: ...ld occur or the Sensor Support Mount may not stick to the site 2 Remove the Sensor Inserter and Sensor Support Mount from their sterile packages Save the package of the Sensor Inserter it contains the Sensor Code number You need to enter this code number into the Receiver at a later step Note Make note of the Sensor Code found on the Sensor Inserter packaging You will need to enter this code after...

Page 8: ...has been removed and the Insertion Buttons are pressed a needle quickly goes just under your skin to place the Sensor Do NOT press the buttons until you are ready to insert the Sensor 7 Hold the black part of the Sensor Inserter avoiding the blue release tabs Press down firmly on both grey Insertion Buttons on the top of the Sensor Inserter Try not to press the Sensor Inserter assembly down into t...

Page 9: ...r serial number the Sensor Code screen is displayed 4 Using the Scroll Wheel enter the 3 digit Sensor Code from the Sensor Inserter packaging CAUTION The code numbers MUST match to ensure accurate glucose results Note Once the Sensor Code has been entered and you have pressed Accept you cannot change the Sensor Code number Be careful If you enter the code incorrectly you have to replace the Sensor...

Page 10: ...rglycaemia GLUCOSE READING Control Solution No Home Low GLUCOSE READING Control Solution No Home High Notes Use a new sterile lancet each time NEVER use a lancet or lancing device for more than one person Follow instructions for your lancing device Apply blood sample and wait for results 1 Gently apply blood to ONLY ONE sample area of the test strip Only a very tiny drop of blood is needed Note Do...

Page 11: ...ting the Transmitter within approximately 1 minute you may get a screen which reads TX DETACHED Transmitter has detached from Sensor CGM is not available Replace the Sensor to continue CGM If this screen is displayed press OK If you don t separate the Transmitter from the Sensor Support Mount the Receiver displays a screen which reads Did you remove the Sensor Select Yes to end CGM Press Yes 3 Dis...

Page 12: ...Data Loss System Beep Pulse Harp Zing Delight Rhythm Vibe vibrate Off Note Exercise caution when turning off alarms For example if you turn off the Low Glucose Alarms you do NOT receive text audible nor vibratory notification for low glucose events Glucose Alarms Settings Type What It Is Settings Available OK LOW GLUCOSE Low glucose below 75 mg dL 74mg dL The low glucose alarm notifies you when yo...

Page 13: ...s for more than 12 hours you must select Menu Alarms Tones and turn off alarms one at a time Day Night Glucose Alarms GLUCOSE ALARMS Night Start 21 00 Day Start 07 00 Next Menu Alarms Glucose Alarms Glucose Alarms Options Comments Day Start 00 00 24 00 Day and Night Alarm settings allow you to define different glucose alarm thresholds for different times of the day Night Start 00 00 24 00 Day and ...

Page 14: ... wide variety of indoor and outdoor activities Depending on the activity there are some considerations to be aware of Alarm Snooze Setup The FreeStyle Navigator II System includes a Low Glucose Snooze and High Glucose Snooze feature that allows you to set how frequently a Low Glucose or High Glucose Alarm repeats after you have already cleared the alarm Accept SNOOZE SETUP Low Glucose 30 min High ...

Page 15: ... 2 Wash the skin in mild soap and water and dry thoroughly 3 Check to make sure the inside of the skin is completely dry before putting it back on the Receiver Travelling Check with local authorities prior to departure as rules and regulations may change without notice Follow these guidelines when travelling Notify security personnel of the presence of the device when going through security system...

Page 16: ...000 feet Operating Pressure 1 0 bar 14 7 psia sea level to 0 7 bar 10 1 psia 3 048 metres 10 000 feet Sensor Life Up to 5 days Sensor Operating Skin Surface Temperature 25 ºC to 40 ºC 77 ºF to 104 ºF Glucose Result Range 20 to 500 mg dL 1 1 to 27 8 mmol L Receiver Memory 60 days of normal use including continuous glucose readings stored every 10 minutes and daily blood glucose readings Transmitter...

Page 17: ...y charge An icon without a plug indicates that the battery is not currently being charged Charging An icon with a plug indicates the battery is charging Charged If the entire icon is green this indicates the Receiver is completely charged Audio Vibrate Audio and Vibrate The audio and vibratory settings for your Receiver The settings apply to glucose data loss and system alarms but not to your Prog...

Page 18: ...facturer s declaration electromagnetic immunity The System is intended for use in the electromagnetic environment specified below The customer or the user of the System should assure that it is used in such an environment IMMUNITY test IEC 60601 test level Compliance Level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Flo...

Page 19: ... by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures object...

Page 20: ...o frequency of transmitter m 150 kHz to 80 MHz 1 2 d 80 MHz to 800 MHz 1 2 d 800 MHz to 2 5 GHz 2 3 d 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the max...

Page 21: ...Abbott Diabetes Care Inc 1360 South Loop Road Alameda CA 94502 USA PRT22683 101 in various jurisdictions 2011 2013 ART22683 101 Rev B 01 15 ART22683 101_rev B indd 37 1 7 15 11 43 AM ...

Reviews: