IMMUNITY
test
IEC 60601
test level
Compliance
Level
Electromagnetic
environment – guidance
Conducted RF
IEC 61000-4-6
6 Vrms
150 kHz to
80 MHz
6 Vrms
N/A
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
10 V/m
N/A
Proximity fields
from RF wireless
communications
equipment IEC
61000-4-3
See table on next
page
Compliance to the
tested levels
Portable RF communications
equipment (including peripherals
such as antenna cables and
external antennas) should be used
no closer than 30 cm (12 inches) to
any part of the System, including
cables specified by Abbott Diabetes
Care. Otherwise degradation of the
System could result.
220