Accuracy of the Reader's built-in meter
Results from the Reader's built-in meter may vary slightly from your
actual blood glucose value. This may be due to slight differences in
technique and natural variation in test technology. The table below
shows the results of a study where 119 typical users used the built-in
meter to check their blood glucose level. For example, in the study, the
built-in meter gave results within 15% of true blood glucose level 115
out of 119 times.
Accuracy results for all glucose concentrations
Difference
range between
the true blood
glucose level
and the built-in
meter result
Within ± 5 mg/dL
and 5%
Within ± 10 mg/dL
and 10%
Within ± 15 mg/dL
and 15%
Within ± 15 mg/dL
and 20%
The number
and percent
that match true
blood glucose
level within X%
68/119 (57.1%)
105/119 (88.2%)
115/119 (96.6%)
116/119 (97.5%)
133