Table 7b: Sensor Survival Rate Over Wear Duration (Pediatric; n=139)
Day of Wear
Number of Sensors
Survival Rate (%)
1
137
98.6
2
136
97.8
3
134
97.1
4
133
96.4
5
133
96.4
6
133
96.4
7
133
96.4
8
131
94.9
9
126
91.3
10
124
89.9
11
122
88.4
12
120
87.0
13
114
83.4
14
104
78.1
209