58
E.24 Test History Export
Error: Test History export failed.
Action: Check that the USB memory stick is connected, formatted correctly
and that it has enough free memory. If the error persists, contact Abbott
Technical Support.
E.25 Operator Limit Reached
Error: Operators cannot be created because the maximum operator limit has
been reached. The maximum number of operators is 1000.
Action: Delete obsolete operators in order to create new operators. If the
error persists, contact Abbott Technical Support.
E.26 Log Export Failed
Error: Log export failed.
Action: Check that the USB memory stick is connected, formatted correctly,
and that it has enough free memory. If the error persists, contact Abbott
Technical Support.
E.27 Last Name Invalid
Error: The patient's last name is invalid.
Action: Tap OK to try again. Last name must be 2-20 characters long. Field
names are case sensitive and must be alpha numeric. If the error persists,
contact Abbott Technical Support.
E.28 Date of Birth Invalid
Error: Date of birth invalid.
Action: Date of birth must contain only numeric characters separated by
"/" and have no spaces. The correct format is mm/dd/yyyy. If error persists,
contact Abbott Technical Support.
E.29 Drawer Open
Error: If a test is being run in Walk Away mode and the user opens the drawer,
this error will appear.
Action: Close drawer to continue.
E.30 Assay Add Error
Error: Assay can not be added.
Action: Check Product Insert to see if the assay is supported for this
instrument. If error persists, contact Abbott Technical Support.