DIGIVAL User Manual
55
E.6 Drawer Open
Error: Drawer is open.
Action: Close the drawer to continue. If the error persists, contact Abbott
Technical Support.
E.7 Test Device ID Invalid
Error: Test device ID invalid.
Action: Check foil pouch to ensure barcode was scanned or manually entered
correctly. Ensure that the test is supported by the current version of the
DIGIVAL. To check software version tap
Home> Preferences> About. The
software version will indicate which test types are supported. Tap
OK to try
again. If the error persists, contact Abbott Technical Support.
E.8 Test Type Not Supported
Error: Test type not supported by current software version.
Action: Check if the software is up to date, or tap OK to try again with a
supported test device. To check software version tap
Home> Preferences>
About. The software version will indicate which test types are supported. If
problem persists contact Abbott Technical Support.
E.9 Test Device Expired
Error: Test device has expired.
Action: Tap OK and try again using an in-date test device. If problem persists
contact Abbott Technical Support.
E.10 Tray Barcode Read Failure
Error: Tray barcode could not be read.
Action: Open drawer, check tray barcode is not obscured and close drawer to
try again. If problem persists contact Abbott Technical Support.
E.11 QC Test Required
Error: QC test required to continue.
Action: The DIGIVAL’s QC policy requires a positive and negative QC test
for this test type and lot number to be completed in order to continue. Please
complete a positive and negative QC test to continue.