6 DIGIVAL
™
Calibration
Home>Calibration Check
A calibration check verifies that the DIGIVAL’s internal camera is functioning
correctly, with a lens free from debris, and that the DIGIVAL is working to
specification.
To ensure proper function, a calibration check of the DIGIVAL must be
performed regularly; at least every 30 days. If a calibration check has not been
done in the last 30 days, an error message displays. No tests can be processed
until the calibration check is up to date.
Two Calibration Check Cards, supplied with an expiry date, are provided for
this purpose. Additional Calibration Check Cards can be ordered. See Section
20, “Ordering and Contact Information.”
To perform calibration:
1.
From the
Home menu, tap
Calibration Check. The
system verifies that there
is enough memory to start
the calibration check and
then guides you through the
calibration check steps:
2.
Open the drawer and remove
the tray.
3.
Place the Calibration Check
Card into its position as shown
on the screen.