DIGIVAL User Manual
40
4.
Scan or enter the Test Device ID.
•
To scan: Place the barcode on
the foil pouch under the barcode
scanner.
•
To enter it manually: Tap
Abc to
access the keypad and then
?123
to access the numeric keypad.
Enter the number printed on the
foil pouch (below the barcode)
and then tap
OK.
5.
Scan Patient ID by placing the
Patient ID barcode under the
barcode scanner, or tap
Abc to
enter it manually using the keypad.
The Patient ID must be
2-20 characters, may contain
alphanumeric and special
characters such as # and *, and
have no spaces. Patient IDs are
case-sensitive. User is prompted
to enter additional patient fields if
configured.
6.
Confirm Patient and Test
information. Check that the
Patient and Test information is
correct and tap
OK to confirm.
If the Patient information is not
correct, tap
Back to return to the
previous screen to reenter.
Caution:
Patient information and Test Type cannot be edited once testing has
begun.
7. Refer to assay Product Insert
for sample preparation and assay
procedure. Insert Test Device at
appropriate time according to
Read
Mode chosen.