DIGIVAL User Manual
27
7.3.4 Assays
Home>Settings>Test Settings>Assays
Use the Assays menu option to configure new assays on the DIGIVAL, as they
become available.
1. Tap
Add Assay
2. Scan the 2D barcode provided with
the new assay kit.
3. The instrument will reboot itself
to complete the new assay
configuration.
Note: "Assay is already up-to-date" will
display on screen if attempt is made to
add an existing assay.
7.3.5 Walk Away Mode
Home>Settings>Test Setting>Walk Away Mode
Use the
Walk Away Mode menu
option to configure patient tests to be
run in
Walk Away Mode. Walk Away
Mode allows the DIGIVAL to time
the test for the user, so that the read
occurs at the time specified in that
assay's product insert.
Note: Not all assays will have this
option. Refer to Product Insert for
details.
Summary of Contents for DIGIVAL LFR-024
Page 1: ...LFR 024 USER MANUAL DIGIVAL...
Page 2: ......
Page 3: ...DIGIVAL User Manual DIGIVAL User Manual...
Page 6: ......
Page 77: ......
Page 78: ......
Page 79: ......