Replacing the EPG Batteries
The EPG uses two nonrechargeable CR2450 coin cell batteries. Abbott Medical recommends replacing the
batteries before starting a new trial and only using the batteries included in the External Pulse Generator,
2 Port Header.
NOTE: Dispose of batteries safely. Do not treat them as household waste. Refer to the guidelines in
Appendix B (page 34) for more information.
To replace the batteries:
1. Disconnect the EPG header or multilead trial cable from the EPG if either is connected. Also, if the EPG
1.
is communicating with the clinician programmer or patient controller, end the session. Refer to the
clinician’s programming or user’s controller manual for instructions.
2. Press and slide the battery compartment panel outward to open the battery compartment and remove
2.
the panel.
13