Connecting the Multilead Trial Cable to the EPG
To connect a multilead trial cable to the EPG:
NOTE: For further instructions, refer to the clinician’s manual for the multilead trial cable.
1. Remove the EPG header cap to expose the EPG stimulator port.
1.
2. Carefully slide the connector end of the multilead trial cable into the EPG stimulator port.
2.
CAUTION:
▪
The EPG is not sterile. When connecting and testing the EPG with the multilead trial cable, keep the
▪
EPG outside of the sterile field.
▪
To avoid delivering jolting stimulation to the patient, turn off stimulation before plugging in the
▪
multilead trial cable.
21
▪
▪