Table 9. Guidance and Manufacturer’s Declaration – Proximity Fields
Proximity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment
Guidance
IEC 61000-4-3
385 MHz: 27 V/m @ 18 Hz pulse
modulation
27 V/m
Recommended minimum
separation distance d = 0.3 m
450 MHz: 28 V/m @ FM
modulation
28 V/m
710 MHz, 745 MHz, 780 MHz:
9 V/m @ 217 Hz pulse
modulation
9 V/m
810 MHz, 870 MHz, 930 MHz:
28 V/m @ 18 Hz pulse
modulation
28 V/m
1720 MHz, 1845 MHz, 1970 MHz:
28 V/m @ 217 Hz pulse
modulation
28 V/m
2450 MHz: 28 V/m @ 217 Hz
pulse modulation
28 V/m
5240 MHz, 5500 MHz, 5785 MHz:
9 V/m @ 217 Hz pulse
modulation
9 V/m