MODULE 10: MAINTENANCE
10-20
AEROSET
®
Customer Training Guide
94858-103 — July 2003
Clean the Reagent Bar Code Reader Window
The reagent Bar Code Readers are located behind each
Reagent Supply
Center
. The Bar Code Readers scan the bar codes using a laser beam
directed through a glass window located at the back of each Reagent
Supply Center. The window protects the reader from any liquid (water or
reagent) that might be splashed by the Reagent Supply Center pipetting
operations.
Any residue impairs the reader from operating at its highest efficiency.
Clean the windows as needed.
Refer to
As Needed Maintenance Procedures
in
Section 9, Service and
Maintenance
of the
AEROSET System Operations Manual
for more
information on cleaning the reagent Bar Code Reader window.
Clean Cuvettes
A cuvette may become dirty due to build-up over time with usage or if
the water is allowed to dry in the cuvettes repeatedly. The Cuvette Wash
Procedure during a run, or during
START UP
or
SHUTDOWN
may not
be able to completely clean the cuvette. If any of the following errors
occur and are isolated to specific cuvettes, it may be necessary to
manually clean the affected cuvette(s).
•
RL%
,
A#0
,
A#1
, or
A#2
Result Error Codes for samples without
increased concentration
• Sample results are low or high
NOTE:
The cuvette number used for measurement is displayed
in the
Reaction Graph
screen.
Refer to
As Needed Maintenance Procedures
in
Section 9, Service and
Maintenance
of the
AEROSET System Operations Manual
for more
information on cleaning the cuvettes.
Instructor will need the
following for the demo:
• Water
• Lint-free towel
Summary of Contents for Aeroset
Page 1: ...Abbott Laboratories Abbott Park IL 60064 94858 103 July 2003 AEROSET Customer Training Guide...
Page 2: ...ii AEROSET Customer Training Guide 94858 103 July 2003 NOTES...
Page 14: ...xiv AEROSET Customer Training Guide 94858 103 July 2003...
Page 56: ...1 42 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 1 SYSTEM TOUR...
Page 90: ...2 34 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 2 BASIC RUN...
Page 152: ...MODULE 4 RUN OPTIONS 4 26 AEROSET Customer Training Guide 94858 103 July 2003...
Page 158: ...MODULE 4 RUN OPTIONS 4 32 AEROSET Customer Training Guide 94858 103 July 2003...
Page 201: ...Review Questions AEROSET Customer Training Guide 5 41 94858 103 July 2003 NOTES...
Page 204: ...MODULE 5 RESULT HANDLING 5 44 AEROSET Customer Training Guide 94858 103 July 2003...
Page 244: ...6 40 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 6 CALIBRATION...
Page 287: ...Review Questions AEROSET Customer Training Guide 7 43 94858 103 July 2003...
Page 290: ...7 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 7 QUALITY CONTROL...
Page 336: ...8 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 8 CONFIGURATION...
Page 337: ...AEROSET Customer Training Guide 9 1 94858 103 July 2003 MODULE 9 ICT UNIT MODULE 9 ICT UNIT...
Page 352: ...9 16 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 9 ICT UNIT...
Page 373: ...As Needed Maintenance AEROSET Customer Training Guide 10 21 94858 103 July 2003 NOTES...
Page 378: ...10 26 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 10 MAINTENANCE...
Page 415: ...AEROSET Customer Training Guide Glossary 1 94858 103 July 2003 GLOSSARY GLOSSARY GLOSSARY...
Page 450: ...Appendix 14 AEROSET Customer Training Guide 94858 103 July 2003 NOTES HAZARDS...
Page 451: ...AEROSET Customer Training Guide Index 1 94858 103 July 2003 INDEX...
Page 456: ...Index 6 AEROSET Customer Training Guide 94858 103 July 2003 NOTES Index...